![Audio Precision Portable One Plus Access Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/audio-precision/portable-one-plus-access/portable-one-plus-access_user-manual_3008598080.webp)
will display the amplitude ofthe selected channel divided by the
amplitude ofthe reference channel.
The upper center section ofthe display shows the amplitude ofthe
reference channel. The upper center soft key selects the display units.
The usual set ofrelative and absolute units are available.
The upper right section ofthe display can show the frequency ofthe
reference channel or the phase of the selected channel relative to the
reference channel. Frequency is displayed in Hz and phase in degrees,
in the range of–180 to +180 degrees.
Pressing INPUT A or INPUT B will select the chosen channel for
measurement, with the other input channel as the reference. Pressing
GEN will use the gen-mon signal for the same channel as the
reference.
The lower left soft key is used to optimize the use of the meters. There
are two meters used for a ratio measurement, and one of the two loses
accuracy below about 8 mV. Therefore, it is necessary to restrict one
ofthe two measured amplitudes to values above 8 mV.
You can choose which amplitude is restricted to higher values using
the lower left soft key. If the display reads ‘MODE 1’, then the channel
used for the denominator value (the ‘downstairs’ value in the fraction)
should not be less than 8 mV. Ifthe display reads ‘MODE 2’, then the
channel used for the numerator value (the ‘upstairs’ value in the
fraction) should not be less than 8 mV. MODE 1 is the same as older
Portables which did not have this feature.
Ifboth measurements are well above 8 mV, then the MODE setting
does not matter. Ifboth are below 8 mV, the reading will not be valid.
Ifthe restricted meter reads below 8 mV, the display will read ‘Low’.
A table summarizing this information is shown on the following page.
&
For more
information on
amplitude units, see
page 4-71.
4
Oper
ati
o
n
Operation
Controlling the Analyzer : Ratio
4-46
Portable One Plus Access User's Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com