In LEVEL function the filters, additional amplification, and selectable
detectors are excluded from the principal meter circuit so that both
meters become identical.
The secondary meter and reading detector have no selectable filters,
less gain than the primary meter (10 mV minimum signal for
fully-specified performance), and true RMS response only. The
measurement ofthe secondary reading detector is usually displayed in
the upper center section ofthe display. It generally shows the level of
the input signal before processing, and its reading may be used to
calculate some relative readings such as THD+N when % or dB units
are selected.
The AC MAINS function works by measuring the input line voltage
through very large resistors connected directly at the power mains
input connector.
The signal monitored by the loudspeaker or headphones is normally
picked off at the selected input channel, following input ranging but
preceding any filtering. An internal jumper may be changed to move
the pickoff point to the end of the principal meter, following all filtering.
6
Diagrams
Block Diagram
DetailedTechnical Diagrams
Portable One Plus Access User's Manual
6-3