Tuning Accuracy
±
2%
Bandpass Response
Q=5 (2-pole)
Accuracy (at f
o
)
±
0.3 dB, 20 Hz to 120 kHz
Residual Noise
10 Hz-5 kHz
≤
0.25
µ
V [-130 dBu]
5 kHz-20 kHz
≤
0.5
µ
V [-124 dBu]
20 kHz-200 kHz
≤
1.5
µ
V [-114 dBu]
THD+N / SINAD Function
Fundamental Range
10 Hz to 100 kHz, THD+N mode
400 Hz or 1 kHz, SINAD mode
Measurement Range
<0.001%-100%
Accuracy
±
1 dB, 20 Hz to 120 kHz harmonics
Measurement Bandwidth
LF -3 dB
<10, 22 or 400 Hz
HF -3 dB
22K, 30k, 80k, or >300 kHz;
option filters are also functional
Residual THD+N
5
25 Hz-20 kHz
≤
(0.0025% + 3.0
µ
V), 80 kHz BW [-92 dB]
10 Hz-50 kHz
≤
(0.010% + 10
µ
V), >300 kHz BW [-80 dB]
Minimum Input
25 mV for specified accuracy in AUTO
TUNE mode, 800
µ
V [-60 dBu] in other
modes.
Nulling Time
Typically 2-3 seconds above 25 Hz,
Increases in a “1/V” fashion for inputs
below 25 mV [-30 dBu]
Note 5
System specification including contribution from generator. Generator
residual THD may limit system performance below 25 Hz if output is
>20.0 Vrms balanced, or 10.0 Vrms unbalanced.