generator monitor (gen-mon) path for the primary measurement. A
second press ofthe GEN key will select the gen-mon path for both the
primary and secondary measurements.
The top center and top left soft keys can be used to select the units for
the level measurement. Either soft key changes the units for both
measurements.
The bottom center soft key allows selection of high-impedance (around
100 k
Ω
) or low-impedance (600
Ω
) input termination for analyzer
channel A. The bottom right soft key allows the same selection for
analyzer channel B.
The Level function provides the best amplitude accuracy of which the
Portable One Plus Access is capable, as long as the measured signal is
above about 10 mV. The Amplitude function provides wider dynamic
range, but at reduced accuracy.
When in Level function, the bargraph panel provides a dual bargraph
showing the level ofboth channels. Similarly, the sweep panel will
measure both channels simultaneously, showing the selected channel
as a solid line and the alternate channel as a dotted line. Cursor
measurements will display the results for both channels.
&
For a more
detailed
discussion of
units, see page
4-71.
&
For more
information on
bargraphs, see
page 4-57.
&
For more
information on
sweeps, see
page 4-59.
4
Oper
ati
o
n
Operation
Controlling the Analyzer : Level
4-24
Portable One Plus Access User's Manual