Accuracy
Analog
7
±2 deg, 10 Hz to 5 kHz
±3 deg, 5 kHz to 20 kHz
±4 deg, 20 kHz to 50 kHz (with performance
option)
Digital
±2 deg, 10 Hz to 0.45
´
SR
Resolution
0.01 deg
Minimum Input (both channels)
Analog
1 mV (S/N >40 dB)
Digital
–60 dBFS (S/N >40 dB)
Residual Noise
(Amplitude Function only)
Analog
A-weighted
£
1.2 µV rms
Dolby 2K
£
2.0 µV rms
IEC 468 (CCIR)
£
6.0 µV Qpk
20 kHz LP
£
1.6 µV rms
Digital
A-weighted
£
–142 dBFS
Dolby 2K
£
–134 dBFS
IEC 468 (CCIR)
£
–127 dBFS
20 kHz LP
£
–140 dBFS
THD+N Function
Fundamental Range
10 Hz to 0.47
´
SR
Measurement Range
0 to 100%
Accuracy
±0.3 dB, 10 Hz to 0.45
´
SR (no filters selected)
Residual THD+N
Analog
£
0.0009% [–101 dB] + 1.6 µV (with 20 kHz LP)
Digital
–138 dBFS
Minimum Input for Counter Tuning
Analog
1 mV (S/N > 40 dB)
Digital
–65 dBFS (S/N > 40 dB)
Notch Tuning Modes
Counter Tuned, Sweep Track, Agen Track, Dgen
Track, or Fixed (set by direct entry)
Amplitude & THD+N Filters
(One filter from each of the following 3 groups may be enabled for 3 filters total)
Low Pass Group
Fs/2 (no BW limiting)
20 kHz (6-pole elliptic, 0.1 dB ripple, 110 dB
stopband)
15 kHz (6-pole elliptic, 0.1 dB ripple, 110 dB
stopband)
User defined (6-pole max)
Chapter 6: Specifications
Getting Started with ATS-2
49
7
Both analyzer input channels must have same coupling (ac or dc) selection. Analog
accuracy is valid for any input signal amplitude ratio up to ±30 dB.
Содержание ATS-2
Страница 1: ...Getting Started with Your ATS 2 ATS 2 ...
Страница 2: ......
Страница 3: ...Getting Started with ATS 2 An Introductory Guide to ATS 2 APIB and GPIB Configurations ...
Страница 8: ...iv Getting Started with ATS 2 ...
Страница 16: ...Chapter 1 General Information Other Documentation for ATS 2 6 Getting Started with ATS 2 ...
Страница 28: ...Chapter 3 Hardware Overview ATS 2 Rear Panel 18 Getting Started with ATS 2 ...
Страница 66: ...Chapter 6 Specifications 56 Getting Started with ATS 2 ...