Use of the CFD-50
• Check the temperature display daily and make sure the chamber temperature is between 2-8 °C before using.
• Do not overfill the vaccine storage chamber. Leave gaps between vials and boxes to enable cold air circulation
within the chamber. Use trays to organize vaccines and keep boxes from touching the walls or sitting directly on
the floor of the storage chamber.
• Keep vaccine chamber door tightly closed and latched when not in use to maintain the storage temperature inside
• There is no user-adjustable thermostat or sensors in the CFD-50.
• Designate a person to inspect and record the daily operation to ensure the safe use of the CFD-50.
• Only approved vaccines or medical supplies should be stored in the refrigerator. Do not store any foods, beverages,
or flammable, explosive, volatile items (such as alcohol or gasoline) or other dangerous items in CFD-50.
5.1
Temperature Display and Status LED
The top display shows the actual temperature in the CFD-50 vaccine storage chamber. The temperature sensor and
thermostat are factory calibrated and set. The user does not need to adjust any temperature settings.
The red LED blinks when an alarm is occurring.
Current temperature and status
①
Status LED
The LED to the left of the display will be:
green while the compressor is running, or
blink red during alarms
②
Current vaccine chamber temperature
Temperature display showing the coldest point in the vaccine chamber