21
Popping sound
Thermal contraction and expansion of
the refrigeration system may cause
occasional popping sound.
Normal phenomenon, no action
required.
Clicking sounds
Occasional clicking sounds may be
generated by relay or other electrical
parts during compressor’s starts and
stops.
Normal phenomenon, no action
required.
9
CFD-50 Parts and equipment list
See repair manual for more information.
10
CFD-50 Specification
Model
CFD-50
Ambient Operating Temperature and Humidity Range
10
℃
- 43
℃
; Relative Humidity ≤ 95%
Vaccine Storage Chamber Temperature Range
2
℃
- 8
℃
Holdover Time @ +43 °C
120-hr (5-day)
Vaccine Storage Capacity
50 L
Input Voltage Range
110 - 230 V
Input Frequency
50 – 60 Hz
Input Power
350W (Peak)
Refrigerant
R134a/315g
External Dimensions
545 W * 655 D * 1588H mm
Net Weight
94 Kg
11
CFD-50 Disposal and Recycle Guideline
□
Remove the door and door gasket when the CFD-50 is no longer in use.
□
Disposal of the CFD-50 must be made according to local appliance disposal and recycling regulation.
Contact local CFD-50 distributor for any questions not covered in this manual.