8
Hints for your safety
Damage to the device
Prior to first starting up, check whether the mains voltage specified on the type plate
matches the local mains voltage.
For mains supply, only use the power cable supplied (or an equivalent one).
The device may only be connected to a properly installed protective contact socket.
A medical insulating transformer with earth leakage monitor or any similar safety
system acc. to EN 60601-1 is required, if several devices are connected over one
common power supply. The transformer must correspond to the power consumption
of all the devices to be connected.
Pay attention to the ambient conditions specified in chapter 9.0.
The ATMOS® Cam 21 / 31 may not be operated with units not complying with the
requirements of standard EN 60601-1 "Medical Electrical Equipment" and EN 60601-1-2
"Electromagnetic compatibility (Medical Electrical Equipment)".
Check proper assignment when assembling country-specific connections:
green/yellow: non-fused earth conductor, blue: neutral conductor, black or brown:
phase
Do only use the recommended accessories and compounds.
Protect the device against direct solar radiation and keep it away from heaters.
The lens of the focusing optic should be protected against mechanical contact.