3.7 Installation and space requirements
NOTE – load at the installation site
If several devices are installed in one place, the static load
capacity of the installation site must be ensured.
Installing the device (Fig. 1):
Place the SunCool either on the floor or on a stable, non-
flammable lab bench and align horizontally.
The liquefier unit of the chiller must be installed at the
following distances from adjacent surfaces regardless of
whether it is installed inside or outdoors (the stands shown
in the bottom of Figure 1 are suggested for mounting in
an outdoor installation):
Fig. 2:
A (to the rear)
15 cm
B (to the front)
60 cm
C (to the left)
15 cm
D (to the right, valve side)
40 cm
E (to the top)
min. 2 m
Weights and dimensions of the chiller:
Liquefier unit:
30 kg
Evaporator unit:
approx. 7 kg
Dimensions (W x D x H in cm):
Liquefier unit
70 x 60 x 35
Evaporator unit:
35 x 32 x 33