13

R

EPLACEMENT

 

OF

 B

ROKEN

 Q

UARTZ

 S

LEEVE

SAFETY EQUIPMENT REQUIRED TO PERFORM TASK

CAUTION: Broken Quartz is SHARP. It is recommended that protective goggles and gloves are worn when handling.

              WARNING: Disconnect power to water puri

fi

 er. Shut off water supply to water puri

fi

 er via inlet and outlet shut 

off valves. Drain chamber by removing drain plug.

Follow the steps in “

Quartz Sleeve Installation, Cleaning or Replacement

” to remove lamp and quartz sleeve.

1.  To prevent damage to the electrical components, it is necessary to separate the ballast housing from the puri

fi

 er chamber.

• On 

Mighty

Pure

®

 Models MP16A, and MP22A, the ballast housing is mounted to the puri

fi

 er chamber using four (4) 

No. 8 x 3/8” long screws. Using a Phillips cross point screwdriver, carefully remove the four (4) screws, from along the 
sides of the ballast housing, and set aside. Separate the housing from the chamber.

• On 

Mighty

Pure

® 

Models MP36C, and MP49C , the ballast housing is mounted to the puri

fi

 er chamber using four (4) 

1/4”-20 x 3/8” long hex head bolts. Using a 7/16” wrench or an adjustable wrench, carefully remove the four (4) bolts 
with the lock and 

fl

 at washers, from along the top of the ballast housing, and set aside. Separate the housing from the 

chamber.

• 

Keep ballast housing and mounting hardware in a clean, dry area.

2.  Unscrew static gland nuts from each end of the chamber. Remove Te

fl

 on

® 

washer and o-ring from both ends of quartz sleeve. 

Te

fl

 on

® 

washer will sometimes remain within the static gland nut. If so, remove Te

fl

 on

®

 washer from static gland nut before 

proceeding.

3. 

CAUTION: 

Carefully remove as much of the broken quartz sleeve as possible, from each end of the chamber.

4.  To remove fragments of quartz sleeve, hold the puri

fi

 er vertically and shake. The quartz fragments will break and drop out 

of the puri

fi

 er through the gland 

fi

 tting. Flush water through chamber being careful to remove all quartz fragments from the 

interior of the chamber.

5. 

Carefully

 discard all pieces of the broken quartz sleeve.

6.  Align the end of the replacement quartz sleeve with the gland 

fi

 tting of the chamber.

7.  Carefully slide the quartz sleeve into the chamber guiding it through the gland 

fi

 tting. 

NOTE: 

Use care to keep the quartz 

sleeve parallel to the the chamber. Angling the quartz sleeve in any direction could result in the breakage of the quartz sleeve.

8.  As the quartz sleeve nears the far end of the chamber, support the quartz sleeve by inserting your 

fi

 nger through the far end 

gland 

fi

 tting and into the quartz sleeve. This will minimize the possibility of breaking the quartz sleeve as it passes through 

the gland 

fi

 tting.

9.  Center the quartz sleeve in the chamber, making sure the quartz sleeve protrudes an equal distance past each threaded gland 

fi

 tting, of the chamber. 

10. Re-install o-rings, Te

fl

 on

®

 washers, and static gland nuts. Be sure o-rings are placed on quartz sleeve before Te

fl

 on

®

 washer. 

Tighten static gland nuts 

fi

 rmly by hand only, 

DO NOT USE HAND TOOLS

. Tightening with hand tools is likely to cause 

quartz sleeve to break. 

11.  Re-assemble ballast housing to puri

fi

 er chamber, using hardware removed in Step 1. 

NOTE:

 When re-assembling ballast 

housing to puri

fi

 er chamber, electrical power cord should exit ballast housing from the end mounted opposite the drain plug.

12.  When all connections are complete, allow water to enter the water puri

fi

 er at a low 

fl

 ow rate until the puri

fi

 er is full. With the 

puri

fi

 er pressurized, it should be checked for leaks.

13. See “

Lamp Installation or Replacement

” section to properly re-install the lamp into the water puri

fi

 e

r. 

     

   

Содержание Mighty Pure MP16A

Страница 1: ...PDLO LQIR XOWUDYLROHW FRP Extensive Product Information Available at ultraviolet com Document No 98 1376C October 2015 2015 Atlantic Ultraviolet Corporation Installation Operation Maintenance Read an...

Страница 2: ...INTENANCE NOTES 22 23 USER ASSISTANCE 24 WARRANTY 24 PATENT NOTICE 24 These instructions generally describe the installation operation and maintenance of the Mighty Pure line of water purifiers Models...

Страница 3: ...operate without proper electrical ground 12 Do not exceed water purifier s maximum rated flow capacity 13 Do not exceed maximum operating pressure of 100 PSI 14 Read and follow all notices and warning...

Страница 4: ...r and flows into the annular space between the quartz sleeve and the chamber wall Suspended microorganisms are exposed to the ultraviolet rays emitted by the germicidal lamp The translucent sight port...

Страница 5: ...should be located in an area where any possible condensation or leakage from the water purifier any purifier accessory and or plumbing will not result in damage to the area surrounding the water purif...

Страница 6: ...cted to the cold water line only 5 Installation requires that a 5 micron sediment filter or finer be installed in line prior to the water purifier The sediment filter will stop or trap large particula...

Страница 7: ...nstalled temporarily for shipping purposes Please follow these instructions if applicable 1 Pull off Easy Off end caps from static gland nut on each end of the water purifier chamber 2 Unscrew static...

Страница 8: ...8 Figure 3 Initial Installation of Quartz Sleeve...

Страница 9: ...lerts the user to any malfunction detected 6 Solenoid Valve Operated in conjunction with the Ultraviolet Monitor Safety Sensor or Time Delay Mechanism this valve prevents water flow through the water...

Страница 10: ...still in operation On purifiers not equipped with the Ultraviolet Monitor lamp operation can be verified by a visible glow through the translucent sight port This provides an indication of lamp operat...

Страница 11: ...to chamber until approximately 2 inches of the lamp is exposed on the opposite end Hold lamp and remove lamp socket 6 Lamp should now be disconnected on both ends Carefully remove lamp from chamber Be...

Страница 12: ...move Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end with your finger so that...

Страница 13: ...of quartz sleeve hold the purifier vertically and shake The quartz fragments will break and drop out of the purifier through the gland fitting Flush water through chamber being careful to remove all...

Страница 14: ...gent and rinse with clean water Stubborn stains usually can be removed with a dilute hydrochloric acid NOTE Follow all manufacturer s instructions and precautions when handling chemicals Once the quar...

Страница 15: ...or fittings Tighten suspect connection or fitting or shut down purifier drain and remove fitting or connection Clean threads reapply thread sealing tape and reinstall Poor purifier performance AND OR...

Страница 16: ...perature 35 F 100 F 35 F 100 F 35 F 100 F 35 F 100 F Quartz Sleeve 1 1 1 1 Drain Plug 1 4 NPT 1 4 NPT 1 4 NPT 1 4 NPT Lamp Out Indicator Translucent Sight Port Translucent Sight Port Translucent Sight...

Страница 17: ...elay Mechanism MP16A through MP49C Quantum Thermal Optimizer MP16A through MP49C Solenoid Valve MP16A 1 2 through MP22A 3 4 MP36C 1 MP49C 1 1 2 MP49C 1 1 2 Sure Flo Flow Control Valve MP16A 3 4 MP22A...

Страница 18: ...18 REPLACEMENT PARTS Mighty Pure MODELS MP16A MP22A Figure 8 Exploded View MP16A MP22A...

Страница 19: ...Quartz Sleeve 1 15 1111A 1 15 1112A 10 Sight Port Plug 1 30 1075 1 30 1075 11 Drain Plug 1 27 1216 1 27 1216 12 Speed Nut 6 32 2 50 1314 2 50 1314 13A Ballast 120v 60Hz 1 10 0137 1 10 0137 13B Ballas...

Страница 20: ...20 Mighty Pure MODELS MP36C MP49C Figure 9 Exploded View MP36C MP49C...

Страница 21: ...465G4 1 25 1465F5 12 Drain Plug 1 27 1216 1 27 1216 13 Screw 1 4 20 x 3 8 long 4 50 1034 4 50 1034 14 Lock Washer 1 4 4 50 1321 4 50 1321 15 Flat Washer 1 4 4 50 1317 4 50 1317 16 Ballast Housing Cove...

Страница 22: ...22 MAINTENANCE NOTES...

Страница 23: ...23 MAINTENANCE NOTES...

Страница 24: ...een made to determine the patent status of applications illustrated or described in this publication Inclusion in this publication of any design or method of use which may be patented is not to be con...

Отзывы: