12. Measurement Procedure 

 

 

 

 

WARNING

 

Should the unit begin to smell abnormally, become overheated or begin 
to smoke, turn it off immediately and unplug from the AC outlet. 
Continued use could increase the risk of fire or failure. 
Contact your ATAGO distributor for servicing information. 

Avoid unauthorized repairs or modifications, as these could increase the 
likelihood of fire, electrical shock or burning hazards. 

If the unit is dropped or suffers some other heavy impact, please contact 
your ATAGO distributor for troubleshooting information. 
Continued use of an instrument which has been dropped could increase 
fire or smoking hazards.

 

 

 

 

 

CAUTION

 

Avoid  measurement  of  samples  that  could  corrode  the  prism  or  prism 
stage. 

Cleaning liquids up to 160˚C can be used for CIP or SIP. The momentary 
difference between the sample liquid temperature and the cleaning liquid 
temperature must be no more than 80˚C. 
When  using  cleaning  liquids  at  temperatures  over  160˚C,  the  power 
switch must be turned off. 

If  a  sample  residue  has  built  up  on  the  prism,  clean  it  immediately  as 
instructed on Page 43 "13Cleaning the Prism."

 

 

Verify the following points: 

  Proper  mounting  of  the  detection  section  on  the  tank  (See  Page  12  "5.Mounting  the  Detection 

Section"). 

  Mounting of the calculation display section (See Page 13 "6.Mounting the Display Section"). 

  Connection  between  the  detection  section  and  the  calculation  display  section  (See  Page  15 

"7.Connecting the Detection Section to the Display Section"). 

  Appropriate  external  output  cable  connection  to  RS-232C,  recorder  output  and  alarm  output  ports 

(See Page 17 "8.External Output"). 

Turn on the unit as instructed in 8. "9 Turning On the Power" (see page 20). 

In addition, turn on and set external output devices, such as recorders. 

The unit will display 0.0 (or 0.0000 in nD mode) for about 5 seconds when first powered on. 

If the unit is powered on without sample in contact with the prism, the error message LLLL.LL (or L.LLLLL 

in nD mode) will display, following 0.0 (or 0.0000 in nD mode). 

Set  measurement  value  display  modes  (Refractive  index  (nD),  Brix,  or  Concentration)  (See  Page  21 

"10.Setting Measurement Value Display Modes").

 

Perform  reference  (zero)  level  adjustment  as  needed  (see  page  24  "11.Setting  Measurement  SET UP 

Mode"). 

 
 

Содержание 3577

Страница 1: ...3577 E01 In line Refractometer PRM TANK100 FER Instruction Manual Cat No 3577 ...

Страница 2: ......

Страница 3: ... 2 and Fig 8 3 17 2 RS 232C Output 18 3 Alarm Output high and low limiter output 19 9 Turning On the Power 20 10 Setting Measurement Value Display Modes 21 1 Common Operation 21 2 Change from nD Mode to Brix 21 3 Change from Brix Mode to nD 22 4 Change from Brix Mode to Conc 22 5 Confirmation of Measurement Values at Each Mode 23 11 Setting Measurement SET UP Mode 24 1 Calibration Procedure 24 2 S...

Страница 4: ... indications and symbols of those items are as follows Understand them first and then read the following pages to use your PRM TANK100α FER correctly Explanation of indications WARNING If this indication is neglected and the instrument is handled incorrectly serious injury or death may result CAUTION If this indication is neglected and the instrument is handled incorrectly injury and or property d...

Страница 5: ...sult in fire electrical shock or burns If the instrument is dropped or is subjected to a strong shock have it inspected by an ATAGO distributor Fire or malfunction may result if the instrument is used CAUTION Keep the detection section except for the wetted sensor part away from water Do not operate the instrument while the connecting ports of the display section is wet The instrument is protected...

Страница 6: ...ed fiberboard box in which it had been delivered Carefully read this instruction manual and fully understand the function and operation of each part of the instrument before use Inspect each part of the instrument for damage or defects that may have occurred in the shipping or manufacturing process Check the necessary operations such as zero setting according to the instruction manual The manufact...

Страница 7: ...ld and pull the plug when disconnecting the power cable from the outlet If the cable is pulled improperly the plug may be broken and may result in fire or electrical shock Ensure the grounding pin is properly connected when connecting the power cable to an AC outlet The disconnecting device of the main unit is the power cable Make sure that there is no obstacle in vicinity of the plug Connecting C...

Страница 8: ...the temperature is extremely low Do not leave the instrument in a damp place Do not place or drop heavy objects on the instrument Use this instrument under the condition where humidity is 80 at 31 C or lower falling linearly to 50 at 40 C Main supply voltage fluctuation should not to exceed 10 the nominal voltage Installation categories Overvoltage Categories ll The pollution degree is 2 according...

Страница 9: ...the supposition that the Refractive Index in a vacuum is 1 and is called the absolute Refractive Index Generally however this index is seldom used 3 Brix scale The PRM TANK100α FER is programmed with the Brix scale based on the Refractive Index of water nD 1 33299 as the reference 0 The Brix scale represents the weight of sucrose expressed by percentage sucrose weight in grams contained in 100 gra...

Страница 10: ...l 1 Inspection certificate 1 N These components are available with PRM TANK100α FER and built to standard specifications 2 Installation m Remove the socket caps from the display section three caps to access the terminals Fig 3 1 Run the unit from a power supply of 100V AC to 240V AC 50 60Hz Install the unit in an ambient temperature range of 5 to 40 C This unit contains precision components Avoid ...

Страница 11: ...he measurement of its refractive index Brix and or concentration Tank mounting part Mount the tank unit on this part Detection section Display section connection terminal Connect the DS 2 cable to this terminal Air cooling fin Dissipates heat to preserve the unit s electronic components when measuring high temperature samples Fig 4 1 Fig 4 2 ...

Страница 12: ...nking digit keys Numerical value change keys Use to change the numerical value in SET UP mode Measurement indicator Indicates the current mode of measurement that is continuous or halted During use the yellow light is blinks SET UP mode indicator Yellow light will be turned on at the SET UP mode H over and L over displays If a high limit or low limit alarm is set in the SET UP mode and then the hi...

Страница 13: ...erminal Use to connect the calculation display section to a personal computer Recorder output terminal Use to connect the calculation display section to a recorder Detection section display section connection terminal Connect the DS 2 cable to this terminal Power cable Insert into a 100V AC to 240V AC outlet Calculation display section rear view Fig 4 4 ...

Страница 14: ...he unit turned on may increase the likelihood of electric shock CAUTION Make sure that the tank is empty before the installation 1 Installation Fig 5 1 Install the unit using a packing FER 3S and clamp band m Position it on the lateral side of the tank When installed on the bottom surface of the tank sump liquid may interfere with measurements Fig 5 1 PRM TANK100α FER Ferrule Prism Packing Tank Ja...

Страница 15: ...rol box The display section is equipped with a wall mount handle on the rear face allowing the part to be mounted to a wall surface Fig 6 1 When mounting the display section on a panel use up to two flat washers to adjust the thickness of the panel as needed and bolt the display section The table assumes panel thicknesses of 1 to 5mm Table 6 1 Fig 6 1 Mounting bolt M6 12 Plain washer Panel surface...

Страница 16: ...Fig 6 2 Table 6 1 Panel thickness Plain washer Schematic 1mm 2pcs 2mm 2pcs 3mm 1pcs 4mm 1pc 5mm 0pc Hexagon bolt M6 12 Plain washer Spring washer 1 9 12 1 6 Accessories Hexagon bolt 2 Plain washer 4 φ11 5 ...

Страница 17: ... OFF and unplug it from the AC outlet prior to attaching or detaching the DS 2 cable to or from the terminal m Remove the socket caps from the display section to access the terminals Fig 7 1 After mounting the detection section on the tank connect it to the display section with the DS 2 cable included Fig 7 2 Fig 7 1 ...

Страница 18: ...16 Fig 7 2 Detection section DS 2 cable for connecting the display and detection sections Back of the calculation display section Front of the calculation display section ...

Страница 19: ...to 160 C are generated as recorder outputs 4 to 20mA by the detection section Insert the recorder output cable included into the port on the rear of the display section See 3 Setting recorder output on Page 27 for SET UP mode instructions Table 8 1 Lead wire color Pin number Cord color Signal name Output 1 Red Refractive Index Brix concentration 4 to 20mA 2 Black Temperature 4 to 20mA 5 to 160 C 3...

Страница 20: ...ble 8 2 Lead wire color Calculation display section D Sub connector PC side Pin number Cord color Signal name Pin number Cord color Signal name 1 Red RXD 2 Red White stripe RXD 2 Red White stripe TXD 3 Red TXD 3 Black GND 5 Black GND 7 RTS 8 CTS Fig 8 4 Fig 8 5 Communications parameters Baud rate 4800bps Data length 8bit Parity ODD Stop bits 1bit Transmitted data format a aaaaa bbb bb cccccc ddd d...

Страница 21: ...ed Low limiter output 3 Black Limiter Com 4 Yellow This pin is not applicable to this unit 5 White This pin is not applicable to this unit 6 Green This pin is not applicable to this unit Fig 8 6 Fig 8 7 High and low limiter output typical applications a Drive a relay b Turn on a LED c Feed a sequencer or the like at the signal level TTL R resistance and V are not necessary if the input to a sequen...

Страница 22: ...repair or replacement cables CAUTION Do not handle plugs with wet hands When disconnecting the power cable from an AC outlet hold it by the plug pulling the cable could damage it increasing the likelihood of fire or electric shock When plugging the power cable to an AC outlet ensure that the ground pin is connected as well Make sure that the DS 2 cable is functioning properly Ensure each output ca...

Страница 23: ...SET UP mode the SCALE key is the cancel key One of the following modes may be selected in SET UP mode 1 bIAS 2 ALAr 3 rEC 4 Conc 5 dP 2 Change from nD Mode to Brix During measurement the READING light blinks and the display will show nD readings Fig 10 1 By pressing the START STOP key the READING light will turn off and StoP will display ceasing measurement Fig 10 2 The nD light will remain illumi...

Страница 24: ...l blink and measurements will be displayed as nD values Fig 10 5 Fig 10 5 4 Change from Brix Mode to Conc During measurement the READING light blinks and the display will show Brix readings Fig 10 4 By pressing the START STOP key the READING light will turn off and StoP will display ceasing measurement Fig 10 2 The Brix light will remain illuminated Press SCALE twice and the Conc light blinks indi...

Страница 25: ...t Press START STOP to stop measurement StoP will be displayed on the measurement value display Press SCALE key once and the last value will be displayed Press SCALE again to display results in the next measurement mode i e nD Brix Conc in turn Pressing SCALE while Conc is displayed will bring up SET UP mode StoP will be displayed if left inactive for 10 seconds ...

Страница 26: ...nstructed in 9 Turning On the Power on page 20 Perform calibration in the appropriate measurement scale for your sample i e nD Brix or Conc The example below is for calibration in Brix measurement mode Press the START STOP key to pause the measurement Press the SCALE key three times The SET UP light will then blink and SEt will be displayed on the measurement value display Fig 11 1 If left inactiv...

Страница 27: ...ition and then use key to set a standard value Fig 11 4 and Fig 11 5 Press the SCALE key to abort the change and return to the previous 1 bIAS screen Fig 11 3 Press START STOP key to reset the factory default setting CL appears on the measurement value screen Fig 11 6 and it returns to the previous 1 bIAS screen After setting a standard value press the ENTER key to start data writing The measureme...

Страница 28: ...g 11 7 When L OVER lights comes on Use the keys to change the value and select the decimal place with the keys Fig 11 8 Press the SCALE key to return to the previous 2 ALAr screen Press the ENTER key to set the selected value The H OVER setting will appear next Adjust the H OVER value in the same manner as in step and press the ENTER key to confirm Fig 11 8 Set the period of time the measurement v...

Страница 29: ...rature display will indicate Lo Fig 11 11 and the screen for setting the recorder output lower limit will appear Use the arrow keys to change the value Press the SCALE key to return to the previous 3 rEC screen Fig 11 11 To confirm the changes press ENTER The temperature display will indicate Hi Fig 11 12 and the screen for setting the recorder output upper limit will appear Use the arrow keys to ...

Страница 30: ...in all cells The instructions below are for programming a user scale using Windows XP Windows 7 and Windows 8 for RS 232C data transmission Windows XP Windows 7 and Windows 8 are registered trade marks of Microsoft Corporation in the United States and other countries 4 1 Items to Prepare Data table the refractive index of the sample and the Concentration according to different temperatures To crea...

Страница 31: ... T3 nD30 nD31 nD32 nD33 nD34 nD35 nD36 T4 nD40 nD41 nD42 nD43 nD44 nD45 nD46 T5 nD50 nD51 nD52 nD53 nD54 nD55 nD56 T6 nD60 nD61 nD62 nD63 nD64 nD65 nD66 C1 C7 Concentration or dilution coefficient etc Enter up to 7 points Fill in any blank cells with 0 0 Input condition C1 C2 C3 C4 C5 C6 C7 C1 C2 C3 C4 C5 C6 C7 T1 T6 Measurement temperature Enter up to 6 points Fill in blank cells if any with 0 0 ...

Страница 32: ...xt measure the other 5 NaOH solutions with the refractometer Change the temperature to 5 0 C 10 0 C 20 0 C 30 0 C and 40 0 C and measure Record the concentration temperature and Refractive Index m For a data table use a measurement range that is wider than the range of the actual sample the PRM TANK100α FER will be measuring Accordingly the user scale in this case has the temperature range of 5 to...

Страница 33: ...0 0 0 0 N 2 0 00 10 00 20 00 30 00 40 00 50 00 0 0 C 1 1 33388 1 36050 1 38500 1 40640 1 42370 1 43590 0 0 C 2 1 33369 1 36010 1 38440 1 40570 1 42280 1 43480 0 0 C 3 1 33299 1 35910 1 38310 1 40410 1 42090 1 43260 0 0 C 4 1 33194 1 35780 1 38160 1 40240 1 41900 1 43050 0 0 C 5 1 33061 1 35640 1 38010 1 40070 1 41710 1 42840 0 0 C 6 0 0 0 0 0 0 0 0 0 0 0 0 0 0 How to read the Table 11 4 T indicate...

Страница 34: ...ting Text File Start up a personal computer installed with Windows Click the Start button at the bottom left of the screen Click program Click Accessories Click Notepad The Notepad should appear Fig 11 13 Fig 11 13 ...

Страница 35: ...A TAB code is automatically adjusted to the width of one character by the computer Represents one character of a line feed code Press the Enter key once If line feed codes are correctly entered the cursor comes to this position Fig 11 14 A Enter the number of measured temperature points B Enter the number of digits after the decimal point for a concentration See the examples of entry C Always ente...

Страница 36: ...t is 30000 data If above the unit cannot output correctly e g in case of 0 000 100 000 limit the range of recorder output to 0 000 30 000 or 20 000 50 000 or reduce the digit number such as 0 00 100 00 to make the data below 30000 Example 0 000 30 000 30000 data 0 000 50 000 50000 data above 1 0000 5 0000 40000 data above 1 000 5 000 4000 data m 2 A displayable value is maximum seven characters in...

Страница 37: ...ata entry is complete Fig 11 16 save the file Fig 11 17 Open the file to make sure that the complete data are saved Click on File and then Exit to close Fig 11 18 A text file has been successfully created Fig 11 18 ...

Страница 38: ... 9 pin D sub connector purchase a USB serial adapter RE 55202 USB serial conversion adapter Download Tera Term from a website such as the one below http ttssh2 sourceforge jp index html en Start Tera Term Select Serial on the New connection dialogbox Select the port number Click OK Check the port number Control Panel System and Security Device Manager Port Click Setup and then Serial port ...

Страница 39: ...tings are conformed to the instrument settings Enter the port number selected in step 1 Click OK Click Setup and then Terminal port Enter the settings as shown below and Click OK Make sure that the Local echo is checked ...

Страница 40: ...1 20 4 6 Transferring Data When preparations for sending computer side and preparations for receiving PRM TANK100α FER side are completed send the created text file NaOH solution according to the following steps Saving received data to a local file Click File and Log and a Tera Term Log window will appear Save with a desired file name Window for Tera Tarm comes out on the screen Select File and cl...

Страница 41: ...our mouse B As a menu screen like this opens specify the hard disk drive storing the user scale created earlier with Notepad Fig 11 22 If you select the hard disk drive the Fig 11 23screen appears Specify the folder storing the user scale created with Notepad Repeat the operation 3 until the file storing a created user scale is updated like Fig 11 24 Fig 11 23 A B ...

Страница 42: ... your mouse Data transfer starts Fig 11 24 Click Open in Fig 11 24 to start the data transfer The instrument will beep once the transfer is completed and then the data as programmed in the PRM TANK100α FER will be displayed on the screen below the original data Compare the two sets of data and make sure that the programming has been executed accurately Fig 11 25 Fig 11 25 4 7 Confirming Data Take ...

Страница 43: ... that the nD is displayed to the fifth decimal place and the Brix is to the second decimal place Press the keys and 4 will appear under the LED light for nD and 1 under the LED light for Brix Fig 11 28 This means that the nD is displayed to the fourth decimal place and the Brix is to the first decimal place 5 for nD and 2 for Brix are displayed when key is pressed 4 for nD and 1 for Brix are displ...

Страница 44: ...e has built up on the prism clean it immediately as instructed on Page 43 13Cleaning the Prism Verify the following points Proper mounting of the detection section on the tank See Page 12 5 Mounting the Detection Section Mounting of the calculation display section See Page 13 6 Mounting the Display Section Connection between the detection section and the calculation display section See Page 15 7 C...

Страница 45: ...pending on the characteristics of the sample solution If the prism is not cleaned after the tank is emptied sample residues may crust on the prism surface The sample flow helps flush out sample residues to some extent Clean in place CIP with hot water or alkaline solutions is a common cleaning method If matter adhering to the surface of the prism cannot be removed by such cleaning it is necessary ...

Страница 46: ...ntration and temperature range H over The alarm setting LED light comes on when the sample measured is above the upper limit Measure samples within the upper limit L over The alarm setting LED light comes on when the sample measured is below the lower limit Measure samples within the lower limit E EEEEE The LED light source has been depleted in refractive index nD mode EEEE EE The LED light source...

Страница 47: ...pendable parts for this unit For their availability contact your ATAGO distributor Part name Part number Summary RS 232C output cable RE 65301 4 pin connector D Sub connector on one end standard 10mm Alarm output cable RE 65305 6 pin connector crimp terminal on one end standard 10mm ...

Страница 48: ...en segment LED Output items Refractive Index nD Brix concentration temperature C Output method DC4 to 20mA RS 232C Alarm output Open collector output for high and low limit settings alarm output Connecting cables Detection section Display section power supply 12V and RS 485 Length standard 15m maximum up to 200m Light source LED D line approximation Material Prism Sapphire Prism stage SUS316L Wate...

Страница 49: ...47 17 Dimensions Detection section Weight 3 3kg Display section Weight 3 3kg 192 240 100 23 10 2 φ6 5 224 180 156 6 224 183 6 208 40 22 5 182 1 0 210 1 0 31 8 ...

Страница 50: ...pened and tampered with the components within the casing the warranty will be invalidated and a charge for repair will be assessed The prism is considered a consumable item Therefore any damage to the prism is not covered under the warranty and is subject to a possible inspection fee ATAGO does not warrant malfunctions resulting from user s actions regardless of the warranty period Any abuse or us...

Страница 51: ...rvice Center nearest you The Authorized Service Center in your area will contact you within 1 to 2 business days North America Europe Canada U S A Mexico Netherlands Italy Germany France Spain U K Belgium Poland Greece Russia Belarus Ukraine Serbia Croatia Romania Central America Middle East Africa Asia Oceania Costa Rica El Salvador Guatemala Panama South America Argentina Bolivia Brazil Colombia...

Страница 52: ...37 2110 customerservice atago usa com TEL 91 22 28544915 40713232 customerservice atago india com TEL 66 21948727 9 66 21171549 customerservice atago thailand com TEL 55 16 3913 8400 customerservice atago brasil com TEL 39 02 36557267 customerservice atago italia com TEL 86 20 38108256 info atago china com TEL 7 812 777 96 96 info atago russia com TEL 234 707 558 1552 atagonigeria atago net TEL 7 ...

Отзывы: