25
⑦
Press the ENTER key.
The SET UP light comes on and "1. bIAS" will
appear on the measurement value display (Fig.
11-3).
※
If left inactive for 10 seconds, the display returns to
the "StoP" screen, following three short error beeps
(Fig. 11-2).
Press the SCALE key to return to the previous "SEt"
screen (Fig. 11-1) (During the SET UP operation, the
SCALE key serves as the cancel key).
Fig. 11-3
⑧
Press the ENTER key to access the SET UP mode.
The most recent measurement value will be
displayed.
※
If the last measurement value was LLLL.LL or
HHHH.HH, the screen for setting the bias will not
appear.
Press the SCALE key to return to the previous "1.
bIAS" screen (Fig. 11-3).
⑨
Use
χ
→
key to move digit position and then use
↑
↓
key to set a standard value (Fig. 11-4 and Fig.
11-5).
※
Press the SCALE key to abort the change and
return to the previous "1. bIAS" screen (Fig. 11-3).
Press START/STOP key to reset the factory default
setting.
"CL" appears on the measurement value screen
(Fig. 11-6), and it returns to the previous "1. bIAS"
screen.
⑩
After setting a standard value, press the ENTER
key to start data writing.
The measurement value display turns off during the
data writing process. This may take a while.
The bias is set and calibration is complete.
The "1. bIAS" screen will be displayed (Fig. 11-3).
⑪
To abort while in the SET UP mode, press the
SCALE key.
"StoP" will appear after 10 seconds of inactivity (Fig.
11-2).
To
start
taking
measurements
with
the
post-calibration data press the START/STOP key.
Fig. 11-4 Example using the Brix display
Fig. 11-5
Example using the Refractive Index display
Fig. 11-6