Press the
key again and the
display turns to [4] (Fig. 16-3).
⑧
Press the
key.
The current upper limit value will blink
on the display (Fig. 16-4).
⑨
Adjust the displayed value to the
desired upper limit value by using the
and/or
keys.
⑩
Press the
key.
The upper limit value is now set and
the display returns to [4] (Fig. 16-3).
⑪
Each time the
key is pressed, the
menu item displayed switches in the
order of: [5] and Brix(%).
Select Brix(%).
N
If 30 seconds pass when in steps
③
through
⑩
above, the
display returns to the continuous
display of the Brix(%).
m
Recorder Output when an error
message is displayed
Please take note of the recorder
output signal when an error
message is displayed on the
main unit.
Fig. 16-3
Example: When the upper limit value of the
recorder output range is set to Brix 60.0%.
Fig. 16-4
Error
Message
Recorder Output
LL.L
4mA
HH.H
20mA
EE.E
20mA
...
Corresponds displayed Brix(%)
value.