10
(3) Installation
CAUTION
◇
Before using the recorder output terminal or the RS-232C output
terminal, remove the two metallic caps by turning them
counterclockwise.
Fig. 3-1
①
Connect the main unit to an AC100-240V main outlet (voltage fluctuations not to
exceed 10%), 50/60Hz with the power cable.
②
The main unit should be supplied with DC24V (allowable fluctuation is ±10%).
The AC adapter AD-32, AD-33 or AD-34 (optional) should be connected to a power
supply of AC 100-240V, 50/60 Hz.
③
The instrument should be installed in a location with an ambient temperature of 5 to
40˚C.
④
Because the instrument incorporates highly precise components, DO NOT install in
locations exposed to direct sunlight or near a heating source, or in an environment
that is dusty or exposed to corrosive gasses.
⑤
The instrument should be installed in a location that is free from vibration. When
installing, take special care to avoid any strong shock to the instrument.
⑥
Do not touch the prism surface with your hand. Finger-prints left on the prism surface
may cause sample to build up on the prism.
In the event of contact with the prism surface, clean carefully with a soft tissue soaked
with ethyl alcohol.
In order to prevent scratches, never clean the prism surface using abrasive materials.