solution is recognised.
(If an error occurs, replace the buffer solution)
9.
Press the key. The value of the buffer solution
that
has been recognised is shown on the display.
10.
Press the
key. The indication
OFFS PH
will
appear on the display. Wait for
30 seconds
for
calibration of the probe.
The instrument will calculate the
OFFSET
value
and
show the quality of the electrode as a percentage
value.
11. After
5 seconds
the indication
CAL PH
will appear
on the display and the
4 pH LED
will flash.
12.
Wash the electrode in
bottle n° 2
containing clean water.
13.
Dip the electrode into
bottle n° 3
containing the
solution with a
pH of 4
leaving it there for
15
seconds
.
14.
Press the
key
15.
The value of the buffer solution that has been
recognised will appear on the display.
(If an error occurs, replace the buffer solution)
16.
If you are using a different buffer solution, correct
the value using the
and keys,
10
17.
Press the
key. The indication
GAIN PH
will
appear
on the display. Wait for
30 seconds.
The instrument will calculate the
GAIN
value and
show the quality of the electrode as a percentage value.
18.
The instrument will save the parameters calculated
and the indication
SAVE PAR
will appear on the
display
for one second. The instrument will then exit the
calibration menu and return to the measurement
mode.
The
HOLD
mode will remain active.
19.
Place the electrode back into the tank, open the
flow
of water again setting it at
60 litres/hour
and press
the
key to exit the
HOLD
mode.
Содержание PC95
Страница 1: ...Manual de Instrucciones PC95 pH cloro temperatura Cod 136317 Manual 0000136317 Versi n r 2 0...
Страница 22: ...Set Point 1 00 ppm HIS 0 10 ppm la regulaci n es la l nea punteada Ton 10 Seg Tof 10 Seg BND 0 30 ppm 22...
Страница 30: ...Instruction Manual PC95 pH chlorine temperature Cod 0000136317 Manual 0000136317 Version r 2 0...
Страница 59: ...Manuale Istruzioni PC95 cod 136317 pH cloro temperatura Manuale 0000136317 Versione r 2 0...