LVB-2 Instruction Sheet
Not rated for measurements within MEASUREMENT CATEGORIES II, III, or IV
N’est pas classé pour les catégories de surtension II, III ou IV
DO NOT TOUCH WHEN TESTING OR AFTER A MALFUNCTION HAS OCCURRED.
NE TOUCHEZ PAS LORS DE L’ESSAI OU APRÈS UN DYSFONCTIONNEMENT DU PRODUIT
CAUTION: Never connect LVB-2 to any mains circuit directly
ATTENTION: Ne jamais connecter directement le LVB-2 à un circuit d’alimentation.
General Information
The LVB-2 is a load box used for the following applications:
• Verifying that the failure detectors for a leakage current test sequence of your Associated Research electrical
safety testing instrument are functioning properly.
• Verifying failure limit thresholds for a leakage current test.
• Verifying the three probe configurations of the instrument (G-L, PH-L and PH-PL) are functioning properly.
The LVB-2 consists of loads based on the 132V/264V test voltages and failure limits listed in the IEC60601-1 test stan-
dard for different types of leakage tests.
Note:
The LVB-2 is not intended to comply with any specific safety agency standard.
Note:
The trip setting may vary up to 10% of the set value based on the combined tolerances of the instrument and
the components used in the LVB-2.
Using the LVB-2
The LVB-2 can be used with the following Associated Research, Inc instrument models:
Model
Capability
OMNIA 2 8206
GB/GC, ACW, DCW, IR, RUN, LLT
OMNIA 2 8207
GB/GC, ACW, DCW, IR, RUN, LLT, Built-in 500VA AC
Source
OMNIA 2 8256
GB/GC, ACW(500VA), DCW, IR, RUN, LLT
OMNIA 2 8257
GB/GC, ACW(500VA), DCW, IR, RUN, LLT, Built-in
500VA AC Source
LINECHEK II 620L
LLT, RUN