© Associated Research 2016 

7

LVB-2 Instruction Sheet

Not rated for measurements within MEASUREMENT CATEGORIES II, III, or IV

N’est pas classé pour les catégories de surtension  II, III ou IV

DO NOT TOUCH WHEN TESTING OR AFTER A MALFUNCTION HAS OCCURRED.

NE TOUCHEZ PAS LORS DE L’ESSAI OU APRÈS UN DYSFONCTIONNEMENT DU PRODUIT

CAUTION: Never connect LVB-2 to any mains circuit directly

ATTENTION: Ne jamais connecter directement le LVB-2 à un circuit d’alimentation. 

General Information

The LVB-2 is a load box used for the following applications:

  •  Verifying that the failure detectors for a leakage current test sequence of your Associated Research electrical 

safety testing instrument are functioning properly.

  • Verifying failure limit thresholds for a leakage current test. 

  • Verifying the three probe configurations of the instrument (G-L, PH-L and PH-PL) are functioning properly. 

The LVB-2 consists of loads based on the 132V/264V test voltages and failure limits listed in the IEC60601-1 test stan-

dard for different types of leakage tests. 

Note: 

The LVB-2 is not intended to comply with any specific safety agency standard.

Note:

 The trip setting may vary up to 10% of the set value based on the combined tolerances of the instrument and 

the components used in the LVB-2.

Using the LVB-2

The LVB-2 can be used with the following Associated Research, Inc instrument models:

Model

Capability

OMNIA 2 8206

GB/GC, ACW, DCW, IR, RUN, LLT

OMNIA 2 8207

GB/GC, ACW, DCW, IR, RUN, LLT, Built-in 500VA AC 

Source

OMNIA 2 8256

GB/GC, ACW(500VA), DCW, IR, RUN, LLT

OMNIA 2 8257

GB/GC, ACW(500VA), DCW, IR, RUN, LLT, Built-in 

500VA AC Source

LINECHEK II 620L

LLT, RUN

Содержание LVB-2

Страница 1: ...LVB 2 Instruction Sheet Leakage Current Verification Box Printed May 10 2016 Item 39669 V1 01 2016 Associated Research Inc 13860 W Laurel Drive Lake Forest IL 60045 USA ...

Страница 2: ...rms to the following Standards Safety EN 61010 1 2010 EMC EN 61326 1 2013 Supplementary Information The product herewith complies with the requirements of the Low Voltage Directive 2014 35 EU and the EMC Directive 2014 30 EU The technical file and other documentation are on file with Associated Research Inc Joseph Guerriero President Associated Research Inc Lake Forest Illinois USA November 20 201...

Страница 3: ... CONFIGURATION SWITCH GND LINE ON OFF by 10A power switch PROBE HI LINE ON OFF by 10A power switch PROBE HI PROBE LO ON OFF by 10A power switch Only one switch can be CLOSED at a time RESISTOR RANGE SPECIFICATION 7 5kΩ 300VAC 20W 2 100PPM 15kΩ 300VAC 10W 2 100PPM 30kΩ 300VAC 5W 2 100PPM 150kΩ 300VAC 2W 2 100PPM 350kΩ 300VAC 0 5W 2 100PPM 1 47MΩ 300VAC 0 25W 2 100PPM 2 97MΩ 300VAC 0 25W 2 100PPM 6 ...

Страница 4: ...n Neutral Closed Reverse Off 2 PASS 350kΩ 1 LLT 132VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 150kΩ 1 LLT 132VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 1 47MΩ 1 LLT 132VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 350kΩ 1 LLT 132VAC 100uA High Limit ...

Страница 5: ...ltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 350kΩ 1 LLT 264VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 2 97MΩ 1 LLT 264VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 1 47MΩ 1 LLT 264VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 6 37MΩ 1 L...

Страница 6: ...ing or caution information to avoid personal injury or damage to the product S il vous plaît se référer au manuel d instructions de mise en garde ou information sur la prudence pour éviter des blessures ou des dommages au produit To indicate hazardous voltages may be present Avertissement des tensions dangereuses qui peuvent être présentes Note pay close attention to the maximum voltage and duty c...

Страница 7: ...ifying failure limit thresholds for a leakage current test Verifying the three probe configurations of the instrument G L PH L and PH PL are functioning properly The LVB 2 consists of loads based on the 132V 264V test voltages and failure limits listed in the IEC60601 1 test stan dard for different types of leakage tests Note The LVB 2 is not intended to comply with any specific safety agency stan...

Страница 8: ...tions table There is no PASS condition for the 7 5kΩ check point There is no FAIL condition for the 6 37MΩ check point The Probe Configuration switch can be set to two different positions depending on the verification type Figure 2 shows the two different settings of the switch This setting should match with the Probe Configuration setting on the Leakage Current test instrument An example of the P...

Страница 9: ... and Ground of the DUT outputs connect to the DUT adapter box 2 DUT Adapter box DUT adapter box for connection of Line Neutral and Ground to the LVB 2 connector terminal 3 L N and G connections This is the socket for the standard IEC320 connector The line cord from the DUT adapt er box connects these ports for making the Line Neutral and Ground connections to the LVB 2 box 4 Probe connections OMNI...

Страница 10: ...ations table and use the rotary switch to select the appropriate Check Point by rotating it clockwise or counter clockwise Ensure that the settings on your electrical safety tester are less than or equal to the maximum recommended voltage current and duty cycle settings of the LVB 2 For each Check Point a PASS test and a FAIL test will be needed to programmed into the leakage test instrument for t...

Отзывы: