![Associated Research LVB-2 Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/associated-research/lvb-2/lvb-2_instruction-sheet_2989482004.webp)
© Associated Research 2016
4
GENERAL
Test Points
Test Condition
Specification
PASS 15kΩ
1. LLT: 132VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 7.5kΩ
1. LLT: 132VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 30kΩ
1. LLT: 132VAC, 6000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 15kΩ
1. LLT: 132VAC, 6000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 150kΩ
1. LLT: 132VAC, 2000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 30kΩ
1. LLT: 132VAC, 2000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 350kΩ
1. LLT: 132VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 150kΩ
1. LLT: 132VAC, 500uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 1.47MΩ
1. LLT: 132VAC, 100uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 350kΩ
1. LLT: 132VAC, 100uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 2.97MΩ
1. LLT: 132VAC, 30uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 1.47MΩ
1. LLT: 132VAC, 30uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
PASS 30kΩ
1. LLT: 264VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%
FAIL 15kΩ
1. LLT: 264VAC, 10,000uA High Limit
2. Maximum Voltage 277VAC
3. Ground = Open, Neutral = Closed, Reverse = Off
2%