MAINTENANCE:

To prevent electric shock do not remove the instrument cover. There are no user serviceable parts inside. Routine 

maintenance or cleaning of internal parts is not necessary. Any external cleaning should be done with a clean dry or 

slightly damp cloth. Avoid the use of cleaning agents or chemicals to prevent damage plastic parts or lettering.

ENTRETIEN:

Pour éviter les chocs électriques ne pas enlever le couvercle de l’instrument. Il n’y a aucune pièce reparable par l’util-

isateur. L’entretien de routine ou le nettoyage des pièces internes ne sont  pas nécessaires. Tout nettoyage externe 

doit être fait avec un chiffon sec ou légèrement humide. Éviter l’utilisation de produits de nettoyage ou des produits 

chimiques pour éviter d’effacer  les  lettres ou d’abimer les pièces en plastique.

OPERATING ENVIRONMENT:

This instrument may be operated in environments with the following limits:

- Indoor Use Only

- Altitude: 2000 m 

- Temperature: 0ºC to 40ºC

- Humidity: Maximum 80% RH at 31ºC decreasing to 50% RH at 40ºC

- Pollution Degree: 2

Symbols Explanation:

Please refer to the instruction manual for specific warning or caution information to avoid personal injury or 

damage to the product.

S’il vous plaît se référer au manuel d’instructions de mise en garde ou     information sur la prudence pour 

éviter des blessures ou des dommages au produit

To indicate hazardous voltages may be present.

Avertissement des tensions dangereuses qui peuvent être présentes

Note: pay close attention to the maximum voltage and duty cycle limitations of each resistor. Applying 

voltages that are higher than the recommended maximum setting or duty cycles greater than indicated 

can cause damage to the LVB-2.

Note: attention à la tension maximale et les limites du cycle de travail de chaque résistance. L’application de tensions 

plus élevées que le réglage maximum recommandé ou cycles de travail supérieures à celles indiquées peut causer 

des dommages à la LVB-2

WARNING

The LVB-2 works with test voltages and currents which can cause harmful or fatal electric shock. To prevent 

accidental injury or death, these safety procedures must be strictly observed when handling and using the 

test instrument.

Les tensions et les courants qui peuvent causer des chocs électriques dangereux ou fatal. Pour éviter les blessures 

accidentelles ou la mort, ces procédures de sécurité doivent être strictement observées lors de la manipulation et 

l’utilisation de l’instrument de test

Содержание LVB-2

Страница 1: ...LVB 2 Instruction Sheet Leakage Current Verification Box Printed May 10 2016 Item 39669 V1 01 2016 Associated Research Inc 13860 W Laurel Drive Lake Forest IL 60045 USA ...

Страница 2: ...rms to the following Standards Safety EN 61010 1 2010 EMC EN 61326 1 2013 Supplementary Information The product herewith complies with the requirements of the Low Voltage Directive 2014 35 EU and the EMC Directive 2014 30 EU The technical file and other documentation are on file with Associated Research Inc Joseph Guerriero President Associated Research Inc Lake Forest Illinois USA November 20 201...

Страница 3: ... CONFIGURATION SWITCH GND LINE ON OFF by 10A power switch PROBE HI LINE ON OFF by 10A power switch PROBE HI PROBE LO ON OFF by 10A power switch Only one switch can be CLOSED at a time RESISTOR RANGE SPECIFICATION 7 5kΩ 300VAC 20W 2 100PPM 15kΩ 300VAC 10W 2 100PPM 30kΩ 300VAC 5W 2 100PPM 150kΩ 300VAC 2W 2 100PPM 350kΩ 300VAC 0 5W 2 100PPM 1 47MΩ 300VAC 0 25W 2 100PPM 2 97MΩ 300VAC 0 25W 2 100PPM 6 ...

Страница 4: ...n Neutral Closed Reverse Off 2 PASS 350kΩ 1 LLT 132VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 150kΩ 1 LLT 132VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 1 47MΩ 1 LLT 132VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 350kΩ 1 LLT 132VAC 100uA High Limit ...

Страница 5: ...ltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 350kΩ 1 LLT 264VAC 500uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 2 97MΩ 1 LLT 264VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 FAIL 1 47MΩ 1 LLT 264VAC 100uA High Limit 2 Maximum Voltage 277VAC 3 Ground Open Neutral Closed Reverse Off 2 PASS 6 37MΩ 1 L...

Страница 6: ...ing or caution information to avoid personal injury or damage to the product S il vous plaît se référer au manuel d instructions de mise en garde ou information sur la prudence pour éviter des blessures ou des dommages au produit To indicate hazardous voltages may be present Avertissement des tensions dangereuses qui peuvent être présentes Note pay close attention to the maximum voltage and duty c...

Страница 7: ...ifying failure limit thresholds for a leakage current test Verifying the three probe configurations of the instrument G L PH L and PH PL are functioning properly The LVB 2 consists of loads based on the 132V 264V test voltages and failure limits listed in the IEC60601 1 test stan dard for different types of leakage tests Note The LVB 2 is not intended to comply with any specific safety agency stan...

Страница 8: ...tions table There is no PASS condition for the 7 5kΩ check point There is no FAIL condition for the 6 37MΩ check point The Probe Configuration switch can be set to two different positions depending on the verification type Figure 2 shows the two different settings of the switch This setting should match with the Probe Configuration setting on the Leakage Current test instrument An example of the P...

Страница 9: ... and Ground of the DUT outputs connect to the DUT adapter box 2 DUT Adapter box DUT adapter box for connection of Line Neutral and Ground to the LVB 2 connector terminal 3 L N and G connections This is the socket for the standard IEC320 connector The line cord from the DUT adapt er box connects these ports for making the Line Neutral and Ground connections to the LVB 2 box 4 Probe connections OMNI...

Страница 10: ...ations table and use the rotary switch to select the appropriate Check Point by rotating it clockwise or counter clockwise Ensure that the settings on your electrical safety tester are less than or equal to the maximum recommended voltage current and duty cycle settings of the LVB 2 For each Check Point a PASS test and a FAIL test will be needed to programmed into the leakage test instrument for t...

Отзывы: