LVB-2 Instruction Sheet
Test Setup
To setup a test, connect the DUT outputs (L, N and G) of the leakage instrument to the line cord connector on the
LVB-2. The two probe connections will also need to be made between the instrument and the LVB-2 for the verifi-
cation of additional leakage test types. Program a test sequence based on the parameters listed on the test box for
each load type. Select the appropriate Probe Configuration on the LVB-2 using the Probe Confguration switch.
Refer to the specifications table and use the rotary switch to select the appropriate Check Point by rotating it
clockwise or counter-clockwise:
Ensure that the settings on your electrical safety tester are less than or equal to the maximum recommended
voltage, current and duty cycle settings of the LVB-2. For each Check Point a PASS test and a FAIL test will be
needed to programmed into the leakage test instrument for the pass/fail verifcation. When all connections have
been made and the appropriate test programmed in the instrument, press the TEST button to begin the test. With
the correct settings entered into the leakage test instrument, the leakage test instrument should record a failure for
the FAIL settings and a pass for the PASS settings.