NMG2
®
Stop-Time Measuring System
Version 3.0.0
www.asm-sensor.com
21
Pos : 42 /Bedienungs anl eitungen/NM G2/Pr ot okoll e/N ac hlauf mess ung - Automatisc he M ess ung @ 6\ mod_1511266374071_78. doc x @ 49842 @ @ 1
Protocol Stop Distance Measurement
–
Automatic Measurement
→
0.
→
1.
→
2.
→
3.
→
4.
→
5.
→
6. / 7. / 8.
→
9.
0. Protocol ID (can be set by the customer, see chapter 4.2 Input of a Protocol ID (machine number) on
page 17).
1. Adjusted measurement start position. At this point the contact will be released and the measurement will
be started.
2. At this velocity standstill will be detected. This velocity can be programmed by the manufacturer between
1 ... 10 mm/s. Thus, very slow machines can also be measured.
Standard adjustment ex-factory is 10 mm/s.
3. Adjusted measuring direction
4.
Position “RELEASE”
5. Parameters to calculate the safety distance ((see page 23) )
6. Stop time
7. Stop distance
8. Velocity at the start of measurement
9. Calculated safety distance.