Picture 5: Attachment lenses and objectives
4.4. Illumination
system
4.4.1. The standard illumination of both mi-
croscopes is the
3W-LED incident light illumina-
tor
. The illuminator can be mount directly on
the microscope body or at an articulated arm.
With help of the articulated arm (also avail-
able as flexible arm) the illuminator can be
moved easily in each position.
The size of the illuminated field can be
changed by rotating the socket of the light col-
lector.
Picture 6: 3W-LED incident light illuminator at
articulated arm
Alternate regarding the 3W-LED incident light
illumination there is also available a 6V/10W
or 6V/20W halogen illumination with the right
transformer.
4.4.2.
LED –ring light illumination
will be
used for different working distances in
combination with or without attachment lenses.
The characteristic of this kind of illumination is
a uniform, shadow free illumination of objects
in all directions.
The assembly will be done with a special
adapter. The adapter has to screw directly
onto the objective of the microscope.
The power connection will be over the built-in
power supply of the large stand or over the
transformer.
Picture 7: LED- ring light illumination
4.4.3. A
transformer
will be needed to use
a 3W-LED incident light illuminator, a LED ring
light illumination, an illuminator 6V/10W, a
illuminator 6V/20W or the dark-field device.
Please note: LED illuminators and halogen illu-
minators need their own transformer. A con-
found of the transformers will caused a dam-
age of the transformer and/or of the illumina-
tor!
Picture 8: Transformer
16