TROUBLESHOOTING:

Unit will not start

Check system’s power connection.

Turn On/Off switch ON for operation.

Check that voltage selector module is set for source 

voltage.
Check fuse. If blown, replace with 6A/125V medium 

time delay fuse for 120V operation, or a 4A/250V slo-

blo fuse for 230V operation.

Unit starts but blows fuse

Check that voltage selector module is set for source 

voltage.
Check vacuum waste container. If full, immediately 

empty to continue operation.
Check vacuum hose assembly for blockage. 
Check that fuse is correct for source voltage.

No water coming out

Check water supply tank for water level and verify 

that cap is tight.
Depress foot switch for 5+ sec. to pressurize supply 

tank.
Ensure the handpiece air control knob is not open 

(turned clockwise) all the way.

Insufficient vacuum

Check vacuum waste container. If full, immediately 

empty to continue operation.
Check vacuum hose assembly for blockage. 
Check that waste container lid is properly seated and 

tightly secured.

Insufficient handpiece operation

Check that handpiece coolant ON/OFF toggle is 

“ON”.
Check that air control valve is turned up adequately.
Check that handpiece tubing is untangled and not 

crimped.
Check handpiece connection for missing gasket.

Compressor makes a chugging sound and does not 

start up properly

Check for sufficient power source and avoid long 

light-grade extension cords (Refer to Figure 8 Exten-

sion Cord Recommendations).

EXTENSION CORD RECOMMENDA-

TIONS:

Use only a 3-wire extension cord that has a 3-blade grounding 

plug. Connect extension cord plug to a matching 3-slot 

receptacle. Ensure that the extension cord is in good condition 

and that its gage wire is the correct size wire to carry the current 

this product will draw. An undersized cord is a potential fire 

hazard and will cause a drop in line voltage, resulting in loss of 

power and causing the product to overheat. Figure 8 indicates 

the correct size cord for the length required and the ampere 

rating listed on the product nameplate. 

If in doubt, use the next 

heavier gage cord. The smaller the gage number, the heavier 

the wire gage.

Disposing of your old product

Each Member State in the European Union (EU) requires that 

the waste from electrical and electronic products be disposed 

of separately from normal household waste. This includes 

electronics, metals and electrical accessories, such as signal 

cables or power cords. Products in this category carry the symbol 

below. 

Contact your local disposal authority and comply with the correct 

disposal method and follow national guidelines for unit disposal.

Содержание ADU-17A

Страница 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Страница 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Страница 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Страница 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Страница 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Страница 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Страница 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Страница 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Страница 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Страница 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420966 Rev G ECO 15498 04 2022...

Отзывы: