Your new Aseptico Express Air Portable Dental System is the 

finest portable electric dental system available. The ADU-17A 

features a handpiece control, 3-way air/water syringe, HVE and 

saliva ejector vacuum systems, a self contained water system, 

oilless air compressor, and vacuum pump. The dual voltage 

system comes in a case with wheels and a handle for maximum 

portability.

Congratulations!

This system is engineered to provide many years of reliable 

service. Please read the instructions provided in this manual 

to receive the best and longest service from your Aseptico 

equipment.

Separate manuals may be provided to cover the operation and 

maintenance of other accessories for your unit.

PACKAGE CONTENTS:

• Portable Case with Compressor and Vacuum Pump
• TA-90D 3-Way Air/Water Syringe
• HVE and Saliva Ejector Vacuum Hoses with Valves
• Water Supply Bottle
• AA-42 Air Foot Control (Complete Assembly with Fittings: PN 

330699)
• Power Cord
• Operation & Maintenance Manual
• Highspeed or Lowspeed Pneumatic Handpiece (Purchased 

separately)
• AWS-1C Cavitron Connector
• Two Spare 6A Fuses for 120V Operation 

 

SAFETY PRECAUTIONS:

Aseptico accepts no liability for direct or consequential injury or 

damage resulting from improper use, arising in particular through 

the non-observance of the operating instructions, or improper 

preparation and maintenance.

WARNING: 

 

Sterilize before first and all uses. Clean, disinfect, 

and sterilize new or repaired instruments before first 

use. Only use sterilized handpieces and instruments 

during treatment. Non-sterile handpieces and instruments 

may cause bacterial or viral infections. Always sterilize 

instruments after operation.

CAUTION:

 

 Always examine unit components for damage before 

commencing treatment. Damaged components must not be 

used and must be replaced.

CAUTION:

 

 For device transport, always keeps the pull handle 

in hand during transport and ensure stable platform before 

releasing.

WARNING: 

 

Use for intended purposes only. Failure to observe 

the operating instructions may result in the patient or user 

suffering serious injury or the the product being damaged, 

possibly beyond repair. Before using this product, make 

sure that you have studied and understood the operating 

instructions.

WARNING: 

 

For use by qualified and trained personnel only.

WARNING: 

 

Do not install where there is a risk of an explosion. 

The Express Air is not intended for operation in the presence 

of flammable anesthetics or gases.

WARNING: 

 

Always operate a high-speed handpiece with water 

coolant. Operating a high-speed handpiece without water 

coolant can cause thermal injury to the patient and damage 

the handpiece.

WARNING: 

 

Only connect device to AC outlet with Protective 

Earth (Safety Ground) connection.

WARNING: 

 

The power supply cord is the mains appliance 

disconnect means.  Before use, position the equipment so it 

is readily accessible.

Expected Service Life:

The expected service life is seven years from the date of first use.

Содержание ADU-17A

Страница 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Страница 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Страница 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Страница 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Страница 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Страница 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Страница 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Страница 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Страница 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Страница 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420966 Rev G ECO 15498 04 2022...