6.

TO INSTALL SALIVA EJECTOR HOSE,

TURN CONNECTOR 1/4 TURN CLOCKWISE

VACUUM LINE 

WITH QUICK 

DISCONNECT

OPERATION FUNCTIONS:

1.  ON/OFF SWITCH: 

Turns the ADU-17A system On or Off.

2.  AIR FOOT CONTROL: 

The handpiece only operates  when the foot pedal is 

depressed. When the foot pedal is not being used, the 

vacuum and syringe remain operational. 

3.  HANDPIECE CONTROLS:

A. 

Off/On Coolant Toggle Switch - Turns Handpiece water 

coolant On/Off.

B.  Air Control Switch - Allows adjustment of air pressure to 

the Handpiece.

C.  Coolant Control Switch - Allows adjustment of water 

coolant to the Handpiece.

4.  THREE-WAY AIR/WATER SYRINGE:

 -

Pressing the left button dispenses water.

 -

Pressing the right button dispenses air.

 -

Pressing both buttons simultaneously dispenses an air/

water mist.

5.  WATER SUPPLY BOTTLE: 

The Express Air incorporates a self contained pressurized 

water system. This system consists of a 1-liter white bottle, 

which dispenses water through the handpiece and 3-Way 

Air/Water Syringe. The Water Supply Bottle threads into the 

Reservoir Connector, located in the upper right corner of the 

case.

To refill the Water Supply Bottle:

A. 

Before removing the Bottle for refilling, unplug the power 

cord from the Express Air unit.

 WARNING:

THE WATER SUPPLY BOTTLE IS PRESSURIZED DURING 

OPERATION. DEPRESSURIZE BOTTLE BEFORE REMOVAL 

BY SLIGHTLY LOOSENING BOTTLE FROM RESERVOIR 

CONNECTOR.

B.  Release pressure from Bottle by slightly loosening bottle 

from reservoir connector approximately 1/8-turn, or by 

depressing the air button on the 3-Way Syringe.

C.  Unscrew and remove Bottle from cap.

D.  Fill Bottle with clean water.

E.  Screw Bottle into the Water Reservoir Connector.

F.  Plug in the power cord.

6.  HVE and SALIVA EJECTOR VACUUMS:

The Express Air is equipped with a 5-liter Waste Tank, 

a High Volume (HVE) hose, and a Low Volume (Saliva 

Ejector) hose. Both High and Low Volume systems operate 

simultaneously from the same vacuum source. To gain 

vacuum on the HVE, open the HVE valve to full open and 

close the Saliva Ejector Valve. To gain vacuum on the Saliva 

Ejector, close the HVE and open the Saliva Ejector to full 

open.

 Fig. 5

Содержание ADU-17A

Страница 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Страница 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Страница 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Страница 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Страница 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Страница 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Страница 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Страница 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Страница 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...