Aseptico ADU-17A Скачать руководство пользователя страница 7

7.

The vacuum Waste Tank contains a ball float shutoff to 

prevent accidental overflow. The Tank should be emptied 

when it is ¾ full.

The High Volume (HVE) vacuum system contains a screen 

inside the Waste Tank container. The screen is used for 

collecting large debris upon vacuum intake. Clean screen 

when necessary.

To empty the Waste Container:

A.    Unplug the power cord.

B.     Disconnect the hoses.

C.     Empty container.

D.     Reconnect hoses.

E.     Plug in power cord.

NOTE:

The waste container has a ball float shutoff to prevent 

accidental overflow. If the waste container becomes full 

and the ball float activates, the vacuum motor may not 

restart, and could potentially blow the fuse. Immediately 

empty waste container to continue operation.

7.  CAVITRON WATER OUTLET:

Provides water via standard 1/4” quick disconnect to 
independent Cavitron or scaler units requiring a water supply.

OPERATION:

After the unit has been set up and you have made yourself 

familiar with the system functions, there are two operating modes 

to select from:

• 

On/Off Switch Operation

 - Switch ‘ON’ activates the 

vacuum and 3-way air/water syringe.

•  Foot Pedal Operation

 - Depressing the Foot Pedal 

activates the handpiece.

1.  Connect the power cord to the Express Air Unit and plug into 

appropriate power source.

IMPORTANT

: Ensure that the Voltage Selector Insert is 

properly configured for the power source - Refer to Page 4.

2. 

FOOT PEDAL OPERATION:

A. 

Turn the On/Off Switch on the front of the Fan Housing 

to the ‘ON’ (I) position. The vacuum and syringe will 

become operational. 

B.  Depress the Foot Pedal: The handpiece will activate. 

The Syringe, HVE, and Saliva Ejector will remain 

operational.

CAUTION:

Some new ADU-17A compressor motors might require 

a 2-week “break-in” when first used. As a precaution 

against stalling the compressor or possibly blowing the 

fuses during this break-in period, it is recommended 

that the foot pedal be removed from the System for 

approximately 5 minutes before each initial daily startup, 

during the first 2 weeks of use.

WARNING:

Always operate a high-speed handpiece with water 

coolant. Operating a high-speed handpiece without 

water coolant can cause thermal injury to the patient 

and damage the handpiece. Consult the handpiece 

manufacturer’s instructions for additional information.

COOLANT 

ON/OFF 

SWITCH

AIR 

CONTROL 

VALVE

COOLANT 

CONTROL 

VALVE

CAVITRON 

CONNECTOR

Fig. 6 - Handpiece Controls

3.  Depress the foot pedal to operate the handpiece. Adjust the 

air pressure and water coolant flow to the desired settings 

(See Fig. 6). (Turn the water coolant control knob clockwise 

Содержание ADU-17A

Страница 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Страница 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Страница 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Страница 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Страница 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Страница 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Страница 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Страница 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Страница 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Страница 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420966 Rev G ECO 15498 04 2022...

Отзывы: