background image

lower than the one set in the parameter 

“F.LF”

 (temperature too

cold).
The relative differential that can be set in parameter 

“F.dF” 

is also

associated with these parameters

.

F.dF

F.dF

tim e

F n

off

O N

F .LF

EP

Tem p.

F .FL

off

off

O N

Remember that the fans functioning can be conditioned by the
“Door open”  function by the digital input.

Notes: It is necessary to pay attention to the correct use of this fans
temperature  control functions because in the typical application of
refrigeration the stop of the fans evaporator stops  thermal  
exchange.

4.10 - ALARM FUNCTIONS

The alarm conditions of the instrument are:
- Probe errors : 

“E1”, “-E1”, “E2”, “-E2”

- temperature alarms:  

“Hi” 

and 

“Lo” 

- External alarm:  

“AL”

- Open door alarm: 

“oP”

The alarm functions of the instrument work on the ALARM led,  on
internal buzzer (if present and programmed by par. “o.bu”) and on
output desired, if configured by the parameters “o1”, “o2”,  
depending on what is set on the said parameters.
Any active alarm is shown on the instrument display with the
lighting up of the ALARM led, the silenced or memorized alarm
status is shown by the  ALARM led flashing .
The buzzer (if “o.bu” = 1 or 3) is activated in alarm and can be
disabled (alarm silencing) manually by pressing any key of the
instrument.
The possible selections of output parameters for the alarm
signalling function are:

= At

 - when one wants  the output to be activated in alarm and can

be disabled (alarm silencing) manually by pressing any key of the
instrument (typical application for sound signal).

= AL

 - when one wants  the output to be activated in alarm status

but cannot be disabled manually and are therefore only disabled
when the alarm status ceases (typical application for a light signal).

= An

 - when one wants  the output to be activated in alarm status

and that they remain activated even when the alarm has ceased
(see par.4.10.4) Disablement (recognition of memorised alarm) can
only be carried out manually by pressing any key when the alarm
has ended (typical application for light signal).

= -At

 - when one wants the function described as At but with an

inverse function (output activated in normal condition and disabled
in alarm status).

= -AL

 - when one wants  the function described as AL but with

inverse logic (output activated in normal conditions and disabled in
alarm status).

= -An

 - when one wants the function described as An but with

inverse working logic (output activated in normal conditions and
disabled in alarm status).
The instrument offers the possibility of arranging the alarm memory
function via the parameter  

“A.tA”.

If "A.tA" = oF, the instrument cancels the alarm signal when the
alarm status ends, if instead it is programmed as "on" , the
instrument maintains the alarm signal when the alarm status ends.
To cancel the alarm memory signal, press any key.
It must be remembered that if an output function is desired with an
alarm memory (=An or =-An) it is necessary to set the parameter
“A.tA” = on.

4.10.1 - TEMPERATURE ALARMS

The temperature alarms work according to the programmed probe
measurement, the type of alarm set in the parameter 

“A.Ay”

 the

alarm thresholds set in parameters 

“A.HA”

 (maximum alarm) and

“A.LA”

 (minimum alarm) and the relative differential 

“A.Ad”.

Through the parameter “A.Ay” it is possible to set if the alarm
thresholds “A.HA” and “A.LA” must be considered as absolute or
relative to the  Set Point , if the reference temperature must be Pr1
or “Au” probe measurement and if the display must be show  the
messages Hi (maximum alarm)/ Lo (minimum alarm) to the
intervention of the alarms or no.
The possible selections of the parameter  “A.Ay” are:
= 1 : Pr1 Absolute Alarms with labels (Hi - Lo)
= 2 : Pr1 Relative Alarms with labels (Hi - Lo)
= 3 : “Au” probe Absolute Alarms with labels (Hi - Lo)
= 4 : “Au” probe Relative Alarms with labels (Hi - Lo)
= 5 : Pr1 Absolute Alarms without labels 
= 6 : Pr1 Relative Alarms without labels 
= 7 : “Au” probe Absolute Alarms without labels
= 8 : “Au” probe Relative Alarms without labels
Using some  parameters  it is also possible to delay the
enablement and the intervention of these alarms.
These parameters are:

“A.PA”

 - is the temperature alarm exclusion time on switching on

the instrument if the instrument is in alarm status when it is
switched on.
If the instrument at power on is not in temperature alarm conditions
the time "A.PA is not considered.

“A.dA”

 - is the temperature alarm exclusion time at the end of

defrosting (and , if programmed, at the end of draining) and at the
end of a continuous cycle.

“A.At”

 - is the temperature alarm delay activation time

The temperature alarm is enabled at the end of exclusion time and
is enabled after the “A.At” time when the temperature measured  by
the probe exceeds or goes below the respective maximum and
minimum alarm thresholds.
The alarm thresholds will be the same as those set on the
parameters “A.HA” and “A.LA” if the alarms are absolute (“A.Ay”=1,
3, 5, 7).

A .A d

A .A d

tim e

A L

o ff

O N

A .L A

H i

A .H A

T e m p .

L o

o ff

o ff

O N

or will be the values  [”SP”+”A.HA”] and [”SP”+”A.LA”] if the alarms
are relative (“A.Ay”=2, 4, 6, 8).

tim e

A .A d

A .A d

A L

o ff

O N

H i

A .LA

S P

A .H A

T em p.

L o

off

o ff

O N

The maximum and minimum temperature alarms can be disabled
by setting the relative parameters "A.HA" and "A.LA" = oF.

4.10.2 -  EXTERNAL ALARM

The instrument can signal an external alarm by activating the digital
input with the function programmed as  “i.Fi” = 4 or 9.
At the same time as the signalling of the configured alarm output,
the instrument visualising 

AL

 and the variable set in parameter

“i.dS”  alternately on the display.

ASCON TECNOLOGIC - Z31Y 

 

- OPERATING INSTRUCTIONS - Vr. 02 - 04/12 -  ISTR-MZ31YENG02 -  PAG. 8

Содержание Z31SY

Страница 1: ...4 10 1 ALARM FUNCTIONS 4 10 EVAPORATOR FANS CONTROL 4 9 DEFROST DISPLAY LOCK 4 8 4 END DEFROST 4 8 3 MANUAL DEFROST 4 8 2 AUTOMATIC DEFROST STARTS 4 8 1 DEFROST CONTROL 4 8 COMPRESSOR PROTECTION FUNCT...

Страница 2: ...status on off off and silenced or memorized flashing 11 Led AUX Indicates AUX output status on on off off or inhibited flashing 12 Led Stand By Indicate the Stand by status 2 PROGRAMMING 2 1 FAST PROG...

Страница 3: ...identifiable by the printed code 103AT 11 visible on the sensor part is compliant with standard EN 13485 Thermometers for measuring the air and product temperature for the transport storage and distr...

Страница 4: ...nt is switched off peak values are always re set However it is also possible to reset these values if the instrument is switched on by using the DOWN key hold for 3 sec during peak visualization The d...

Страница 5: ...put will be on in normal mode Set Point SP active and off in economy mode operation Set Point SP2 active 4 Internal Light output managed by digital input This output will be on when door is opened i F...

Страница 6: ...lts that currently used in the refrigerators systems ct intervals with counts only the compressor function time output ot switched on Mode typically used in the positive temperature refrigerators syst...

Страница 7: ...1 differential Hysteresis The active defrost is shown on the instrument display with the lighting up of the DEFROST led At the end of defrosting it is possible to delay the new start up of the compre...

Страница 8: ...s the alarm signal when the alarm status ends To cancel the alarm memory signal press any key It must be remembered that if an output function is desired with an alarm memory An or An it is necessary...

Страница 9: ...cessary that the device or instrument are being supplied USB SUPPLY AC SUPPLY SUPPLY ADAPTER 12 VDC USB For additional info please have a look at the A01 instruction manual 4 12 2 TVRY REMOTE DISPLAY...

Страница 10: ...obe Calibration i C2 10 oF oF 0 01 9 59 min sec Fan delay after defrost F Fd 37 oF oF on Fan status during defrost F FE 36 1 0 0 0 30 0 C F Differential fan control F dF 35 99 9 99 9 999 C F Low tempe...

Страница 11: ...rs relative to compressor protection and power on delay 99 5 min sec x10 1 0 255 MODBUS Station address for serial communication t AS 60 oF oF 999 Access Password to parameter functions t PP 59 oF oF...

Страница 12: ...21 990 25 C or NTC 103AT 2 10K 25 C 1 digital input for free voltage contacts alternative to Pr2 input Output s up to 2 relay outputs 5 A Res 6 4 A 8 3 A Out2 SPDT 8A 1 2HP 250V 1 3HP 125 VAC 12 A Res...

Страница 13: ...1000 4 4 2KV supply and relay outputs 1KV inputs EN61000 4 5 supply 2KV com mode 1 KV diff mode EN61000 4 6 3V Directive 2006 95 CE EN 60730 1 EN 60730 2 9 Regulation 37 2005 CE EN13485 air S A 2 50 C...

Отзывы: