Ascon tecnologic Y33 Скачать руководство пользователя страница 9

Example  of  electric  defrost  with  evaporator  temperature  control:
The  defrost  end  after  "d.dE"  programmed  time.  During  defrost  the
“dF”  output  switch  on/off  to  control  evaporator  temperature    in
heating mode with  set point “d.tE” and 1° differential (Hysteresis).

The  active  defrost  is  shown  on  the  instrument  display  with  the
lighting up of the DEFROST led
At  the end of  defrosting, it is  possible to delay  the new start  up of
the compressor (output “ot”) at the time set in parameter 

“d.td” 

to

allow the evaporator to drain.
During  this  delay,  the  led  Defrost  flashes  to  indicate  the  draining
state.

4.8.4  -  DEFROSTS  IN  EVENT  OF  EVAPORATOR  PROBE
ERROR

In  event  of  evaporator  probe  error  the  defrosts  occur  at  intervals

"d.Ei"

 and duration 

"d.EE”.

In case an error occurs when the time  remaining to the start or the
end  of  defrost  it’s  lower  than  that  normally  set  the  parameters
related  to  error  conditions  probe,  the  start  or  the  end  take  place
with the shortest time.
The  functions are provided because  when the evaporator probe is
used  the  defrost  endurance  time  is  usually  set  longer  than
necessary  (the  time  “d.dE”  is  a  security  time-out)  and  in  case  is
used the "Dynamic Intervals Defrost System” the interval is usually
set  more  longer  than  what  is  normally  programmed  into
instruments that do not have the function.

4.8.5 - DEFROST DISPLAY LOCK

Through par. 

“d.dL” 

and

 “A.dA”

 it’s possible to define the display

behaviour during defrost.
The  “d.dL”  parameter  pemits  the  display  visualization  lock  on  the
last  Pr1  emperature  reading  (“d.dL”  =  on)  during  all  the  defrost
cycle  until, at the end  of defrost, the temperature has  not reached
the  lock  value  or  the  value  [”SP”  +  “r.d”]  or  is  elapsed  the  time
setted on par. "A.dA". 
Or  it  permits  only  the  visualization  of    label 

“dEF”

  (“d.dL”  =  Lb)

during the defrost cycle and, after the defrost, of label 

“PdF”

 until,  

at the end of defrost, the Pr1 temperature has not reached the lock
value  or  the  value  [”SP”  +  “r.d”]  or  is  elapsed  the  time  setted  on
par. "A.dA".
The  display  will  otherwise  (“d.dL”=  oF)    continue  to  visualize  the
Pr1 temperature measured by the probe during the defrost cycle.

4.9 - EVAPORATOR FANS CONTROL

The control of the fans on the output configured as “Fn” depending
on  determined  control  statuses  of  the  instrument  and  the
temperature measured by the evaporator probe (EP).
In  the  case  that  the  evaporator  probe  is  not  used  or  in  error  ,  the
output  Fn  is  activated  only  depending  on  the  parameters  “F.tn”,
“F.tF” and “F.FE”.
The parameters

 “F.tn”

 e 

“F.tF”

 decides the funs functioning when

the output configured as “ot” (compressor) is off.
When output “ot” is off , it is possible to set the instrument so that
that  the  output  “Fn”continues  to  work  in  cycles  according  to  the
times  programmed  in  the  parameter   

“F.tn”

  (fan  activation  time)

and 

“F.tF” 

(fan deactivation time).

When output “ot” is switched off the instrument activates the output
“Fn for the time “F.tn”, then deactivates it for the time “F.tF” and so
on whilst the otuput “ot” remains off.
Programming  “F.tn”  =  oF  the  output  “Fn”  in  “ot”  off  condition  will
remain switched off.
Programming instead “F.tn” to any value and “F.tF” = oF the output
“Fn” in “ot” off condition will remain switched on.
The  parameter 

“F.FE

”  instead  decides  whether  the  fans  must

always  be  switched  on  independently  of  the  defrosting  status
(“F.FE”=on) or switched off during defrosting (“F.FE”=oF).
In this later case, it is possible to delay the start up of the fans even
after  the  end  of  the  defrosting  of  the  time  set  in  the  parameter

“F.Fd”.

When this delay is active the led FAN flashing to signal the delay in
progress.
When  the  evaporator  probe  is  used  the  fans,  as  well  as  being
conditioned  by  the  parameters  “F.tn”,  “F.tF  and  “F.FE”,  are  also
conditioned by a temperature control.

It  is  possible  to  set  the  disablement  of  the  fans  when  the
temperature  measured  by  the  evaporator probe  is  higher  than  the
one set in the parameter 

“F.FL”

 (temperature too hot) or when it is

lower  than  the  one  set  in  the  parameter 

“F.LF”

  (temperature  too

cold).
The relative differential that can be set in parameter 

“F.dF” 

is also

associated with these parameters

.

Remember  that  the  fans  functioning  can  be  conditioned  by  the
“Door open”  function by the digital input.

F.d F

F.d F

tim e

Fn

o ff

O N

F .LF

EP

T e m p .

F .FL

off

off

O N

Notes: It is necessary to pay attention to the correct use of this fans
temperature  control functions because in the typical application of
refrigeration the stop of the fans evaporator stops  thermal  
exchange.

4.10 - ALARM FUNCTIONS

The alarm conditions of the instrument are:
- Probe errors : 

“E1”, “-E1”, “E2”, “-E2”, “E3”, “-E3”,

- temperature alarms:  

“Hi” 

and 

“Lo” 

- External alarm:  

“AL”

- Open door alarm: 

“oP”

The alarm functions of the instrument work on the ALARM led,  on
internal buzzer (if  present and programmed by par. “o.bu”) and on
output  desired,  if  configured  by  the  parameters  “o.o1”,  “o.o2”  or
“o.o3”,  depending on what is set on the said parameters.
Any  active  alarm  is  shown  on  the  instrument  display  with  the
lighting  up  of  the  ALARM  led,  the  silenced  or  memorized  alarm
status is shown by the  ALARM led flashing .
The  buzzer  (if  “o.bu”  =  1  or  3)  is  activated  in  alarm  and  can  be
disabled  (alarm  silencing)  manually  by  pressing  any  key  of  the
instrument.
The  possible  selections  of  output  parameters  for  the  alarm
signalling function are:

= At

 - when one wants  the output to be activated in alarm and can

be  disabled (alarm  silencing)  manually by  pressing any  key  of the
instrument (typical application for sound signal).

= AL

 - when one wants  the output to be activated in alarm status

but  cannot  be  disabled  manually  and  are  therefore  only  disabled
when the alarm status ceases (typical application for a light signal).

= An

 - when one wants  the output to be activated in alarm status

and  that  they  remain  activated  even  when  the  alarm  has  ceased
(see par.4.10.4) Disablement (recognition of memorised alarm) can
only  be  carried  out  manually  by  pressing  any  key  when  the  alarm
has ended (typical application for light signal).

=  -t

  -  when  one  wants  the  function  described  as  At  but  with  an

inverse function (output activated in normal condition and disabled
in alarm status).

=  -L

  -  when  one  wants    the  function  described  as  AL  but  with

inverse logic (output activated in normal conditions and disabled in
alarm status).

=  -n

  -  when  one  wants  the  function  described  as  An  but  with

inverse  working  logic  (output  activated  in  normal  conditions  and
disabled in alarm status).
The instrument offers the possibility of arranging the alarm memory
function via the parameter  

“A.tA”.

If  "A.tA"  =  oF,  the  instrument  cancels  the  alarm  signal  when  the
alarm  status  ends,  if  instead  it  is  programmed  as  "on"  ,  the
instrument maintains the alarm signal when the alarm status ends.
To cancel the alarm memory signal, press any key.
It must be remembered that if an output function is desired with an
alarm  memory  (=An  or  =-An)  it  is  necessary  to  set  the  parameter
“A.tA” = on.

ASCON TECNOLOGIC - Y33 

 

- OPERATING INSTRUCTIONS - Vr. 01 - 04/16 -  ISTR-MY33-ENG01 -  PAG. 9

Содержание Y33

Страница 1: ...EVAPORATOR FANS CONTROL 4 9 DEFROST DISPLAY LOCK 4 8 5 DEFROSTS IN EVENT OF EVAPORATOR PROBE ERROR 4 8 4 END DEFROST 4 8 3 MANUAL DEFROST 4 8 2 AUTOMATIC DEFROST STARTS 4 8 1 DEFROST CONTROL 4 8 COMPRESSOR PROTECTION FUNCTION AND DELAY AT POWER ON 4 7 TEMPERATURE CONTROL 4 6 OUTPUTS AND BUZZER CONFIGURATION 4 5 DIGITAL INPUT 4 4 MEASURING AND DISPLAY 4 3 NORMAL MODE ECONOMIC MODE AND TURBO MODE 4 ...

Страница 2: ...d silenced or memorized flashing 11 Led AUX Indicates AUX output status on on off off or inhibited flashing 12 CLOCK LED Indicates that the internal clock is activated When on indicates that the current time is presumably correct If blinking indicates that there was a supply black out and therefore the current time may not be correct 13 Led Stand By Indicate the Stand by status 2 PROGRAMMING 2 1 F...

Страница 3: ...hich will be possible to have access to the protected parameters 2 5 RESET PARAMETERS TO DEFAULT VALUE LEVEL The instrument allows the reset of the parameters to values programmed in factory as default To restore to the values of default the parameters set the value 48 to r P password request Once confirmed the password with the key P the display it shows for 2 sec therefore the instrument effects...

Страница 4: ...he controller uses the control functions STAND BY means that the controller does not use any control function and the display is turned off except for the Stand by led If there is no power and then power returns the system always sets itself in the condition it was in before the black out The ON Stand by function can be selected Pressing the key U for at least 1 sec if the parameter t UF 3 Pressin...

Страница 5: ...e Pr1 P1 the measurement of the probe Pr2 P2 the measurement of the probe Pr3 P3 the active set point value SP the label Eco when the instrument is in Eco mode Ec or it can have the numerical display switched off oF Through the parameter i CU it is possible to program an measu re offset that will be applied to the temperature show on the display only if i dS P1 P2 P3 Ec The normal visualisation on...

Страница 6: ...figured t UF or t Fb 1 These commands have a bi stable function Which means that when first pressed the output key is activated while the second is disabled In this mode the AUX output can be turned off automatically after a certain time that can be set on the parameter o tu With o tu oF the output is activated and deactivated only manually using the key U or DOWN AUX or via the digital input Diff...

Страница 7: ...off ON ON Pr1 ot Out During the output inhibition the led OUT Cool o Heat blinking It is also possible to prevent activation of the output after the instrument is turned on for the time set in the parameter P od During the power on delay phase the display shows the indication od alternating with the normal visualisation All the functions are disabled by relative parameters oF 4 8 DEFROST CONTROL T...

Страница 8: ... of the temperature difference between cell and evaporator Every variation of the value of the Active Set Point of the differential r d the start of a continuous cycle or the a defrost execution delete this reference value and any reduction will be performed until the acquisition of a new reference value Defrost by evaporator temperature The instrument starts a defrost cycle when the evaporator te...

Страница 9: ...ssible to delay the start up of the fans even after the end of the defrosting of the time set in the parameter F Fd When this delay is active the led FAN flashing to signal the delay in progress When the evaporator probe is used the fans as well as being conditioned by the parameters F tn F tF and F FE are also conditioned by a temperature control It is possible to set the disablement of the fans ...

Страница 10: ...nstrument visualising AL and the variable set in parameter i dS alternately on the display In alarm conditions with i Fi 5 all the control outputs will be off 4 10 3 OPEN DOOR ALARM The instrument can signal an open door alarm by activating the digital input with the function programmed as i Fi 1 2 or 3 When the digital input is activated the instrument show oP and after the delay programmed in pa...

Страница 11: ...99 9 S HS Minimum Set Point S LS 1 S parameters relative to Set Point Note Def Range Description Par oF oF 0 01 9 59 min sec 99 5 min sec x10 Ritardo compressore dopo sbrinamento sgocciolamento d td 34 20 0 oF 0 01 9 59 min sec 99 5 min sec x10 Lenght max of defrost cycle d dE 33 oF oF 0 01 9 59 hrs min 99 5 hrs min x10 Delay start Defrost by continuous compressor running time d cd 32 oF oF on Lb ...

Страница 12: ...e cycle no without compr output condictioning Et Electrical heating with evaporator tempe rature control d dt 35 3 oF 1 2 3 Buzzer function mode o bu 65 Fn oF ot dF Fn Au At AL An t L n on HE OUT3 function see o o1 o o3 64 dF oF ot dF Fn Au At AL An t L n on HE OUT2 function see o o1 o o2 63 ot oF ot dF Fn Au At AL An t L n on HE OUT1 function oF No function ot Temperature control compressor dF De...

Страница 13: ...ring flaws or faulty material that are found within 12 months from delivery date The guarantee is limited to repairs or to the replacement of the instrument The eventual opening of the housing the violation of the instrument or the improper use and installation of the product will bring about the immediate withdrawal of the warranty s effects In the event of a faulty instrument either within the p...

Страница 14: ...acy 0 5 fs 1 digit Sampling rate 130 ms Display 3 Digit Red Blue optional h 15 5 mm Software class and structure Class A Compliance Directive 2004 108 CE EN55022 class B EN61000 4 2 8KV air 4KV cont EN61000 4 3 10V m EN61000 4 4 2KV supply and relay outputs 1KV inputs EN61000 4 5 supply 2KV com mode 1 KV diff mode EN61000 4 6 3V Directive 2006 95 CE EN 60730 1 EN 60730 2 9 Regulation 37 2005 CE EN...

Отзывы: