Ascon tecnologic TLK 33 Скачать руководство пользователя страница 5

ASCON TECNOLOGIC - TLK33 - OPERATING INSTRUCTIONS -

 PAG. 5

If instead, it is desired that the offset set should not be constant for 
all  the  measurements,  it  is  possible  to  operate  the  calibration  on 
any two points.  
In  this  case,  in  order  to  decide  which  values  to  program  on  par. 
“OFSt” and “rot”, the following formulae must be applied : 

“rot” = (D2-D1) / (M2-M1)  

“OFSt” = D2 - (“rot” x M2)

 

where: 
M1 =measured value 1 
D1 = visualisation value when the instrument measures M1 
M2 =measured value 2 
D2 = visualisation value when the instrument measures M2 
It then follows that the instrument will visualise : 

DV = MV x “rot” + “OFSt”

 

where:  DV = visualised value  

MV= measured value 

Example  1:  It  is  desired  that  the  instrument  visualises  the  value 
effectively  measured  at  20°  but  that,  at  200°,  it  visualises  a  value 
lower than 10° (190°).     
Therefore :  M1=20 ;  D1=20 ;  M2=200 ;  D2=190 
“rot” = (190 - 20) / (200 - 20) = 0,944 
“OFSt” = 190 - (0,944 x 200) = 1,2

 

Example 2: It is desired that the instrument visualises 10° whilst the 
value actually measured is 0°, but, at 500° it visualises a 50° higher 
value (550°). 
Therefore :  M1=0 ;  D1=10 ;  M2=500 ;  D2=550 
“rot” = (550 - 10) / (500 - 0) = 1,08 
“OFSt” = 550 - (1,08 x 500) = 10

 

By  using  par.

  “FiL” 

it  is  possible  to

 

program  time  constant  of  the 

software  filter  for  the  input  value  measured,  in  order  to  reduce 
noise sensitivity (increasing the time of reading). 
In  case  of  measurement  error,  the  instrument  supplies  the  power 
as programmed on par.  

“OPE”. 

This power will be calculated according to cycle time programmed 
for the PID controller. 
By  using  par. 

“InE”

  it  is  also  possible  to  decide  the  conditions  of 

the  input  error,  allowing  the  instrument  to  give  the  power 
programmed on par. “OPE” as output. 
The possibilities of par. “InE” are : 
= Or : the condition occurs in case of over-range or probe breakage 
=  Ur  :  the  condition  occurs  in  case  of  under-range  or  probe 
breakage 
=  Our  :  the  condition  occurs  in  case of over-range or under-range 
or probe breakage 
Using par. 

“diSP”, 

located in the group 

]

PAn”, 

it is possible to set

normal  visualization  of  the  display  which  can  be  the  process 
variable (dEF), the control power (Pou), the active Set Point (SP.F) 
the  Set  Point  operating  when  there  are  active  ramps  (SP.o)  or 
alarm threshold AL1, AL2  (AL1, AL2).

 

Again in the group 

]

PAn”

 the  par. 

“AdE”

 is present that defines

the 3 led shift index functioning. 
The lighting up of the green led = indicates that the process value 
is within the range  [SP+AdE ... SP-AdE], the lighting up of the led – 
indicates  that  the  process  value  is  lower  than  [SP-AdE]  and  the 
lighting  up  of  the  led  +  indicates  that  the  process  value  is  higher 
than [SP+AdE]. 

4.2 - DOUBLE ACTION PID CONTROL 

All  the  parameters  referring  to  PID  control  are  contained  in  the 
group 

]

rEG”.

The Double Action PID control is used to control plants where there 
is  an  element  which  causes  a  positive  increase  (ex.  Heating)  and 
an element which causes a negative increase (ex. Cooling). 
The double action PID controller is used to control plants equipped 
with  an  actuator  that  causes  both  a  positive  increment  (e.g. 
Heating) and a negative increment (e.g. Cooling). 
This  is  the  case  of  temperature  control  using  a  PELTIER  device 
where  it  is  possible  to  obtain  a  heating  effect  polarizing  in  one 
direction and a cooling effect when polarizing in the other direction. 
The  Double  Action  PID  control  works  on  the  output  depending  on 
the  active  Set  Point 

“SP” 

and  on  the  instrument’s  PID  algorithm 

with two degrees of freedom. 
The Double Action PID control algorithm needs the programming of 
the following parameters : 

"Pb"

 - Proportional Band 

"tcr1"

 - Cycle time for HEAT action 

“tcr 2”

 - Cycle time for COOL action 

"Int"

 - Integral Time 

"rS"

 - Manual Reset (if “Int =0 only) 

"dEr"

 - Derivative Time 

“FuOC” - 

Fuzzy Overshoot Control

 

"Prat"

  -  Power  Ratio  or  relation  between  Cooling  power  and 

Heating power  
The connection of the Peltier device to the instrument indicated on 
the wiring diagram always foresees a double action control, as it is 
the  instrument  that  automatically  commands  and  polarizes  the 
Peltier device depending on the action to be effected  (HEATING or 
COOLING). 
In  the  case  that  the  actuator  must  carry  out  only  one  action,  the 
Peltier device can be connected to the instrument in such a way to 
command only activation and not polarization. 

1 8

1 6

1 7

1 5

1 4

+ H E A T

+ C O O L

( A U T O )

+

H E A T

C O O L

+

O N L Y

O N L Y

C O O L

H E A T   &

4.3 - AUTOTUNING

 

AND SELFTUNING FUNCTIONS

 

All  the  parameters  referring  to  the  AUTO-TUNING  and  SELF-
TUNING functions are contained in the group 

]

rEG”.

The  AUTO-TUNING  and  SELF-TUNING  functions  permit  the 
automatic tuning of the PID controller. 
The 

AUTO-TUNING

  function  permits  the  calculation  of  thePID 

parameters by means of a FAST type tuning cycle and, at the end 
of  this  operation,  the  parameters  are  stored  into  the  instrument’s 
memory and remain constant during control. 
The 

SELF-TUNING

 function (rule based "TUNE-IN") instead allows 

control monitoring and the continuous calculation of the parameters 
during control. 
Both functions automatically calculate the following parameters : 

"Pb"

 - Proportional Band 

"tcr1"

 - Cycle time for HEAT action 

“tcr 2”

 - Cycle time for COOL action 

"Int"

 - Integral Time 

"dEr"

 - Derivative Time 

“FuOC” - 

Fuzzy Overshoot Control

 

"Prat"

  -  Power  Ratio  or  relation  between  Cooling  power  and 

Heating power 

To activate the AUTO-TUNING function proceed as follows : 
1) Program and activate the desired Set Point.
2) Program par. 

"Auto"

 as:

- "1”  –  if  auto-tuning  is  desired  automatically,  each  time  the
instrument is switched on, on the condition that the process value is
lower  (with  “Func”  =HEAt)  than  [SP-  |SP/2|]  or  higher  (with  “Func”
=CooL) than [SP+ |SP/2|].
- "2"  –  if  auto-tuning  is  desired  automatically,  the  next  time  the
instrument is switched on, on the condition that the process value is
lower  (with  “Func”  =HEAt)  than  [SP-  |SP/2|]  or  higher  (with  “Func”
=CooL) than [SP+ |SP/2|], and once the tuning is finished, the par.
“Auto” is automatically swapped to the OFF state
- "3"  - if manual auto-tuning is desired, by selecting  par. “tunE” in
the  main  menu  or  by  correctly  programming  key  “U”  as  “USrb”  =
tunE.  The  Autotuning  will  start  at  the  condition  that  the  process
value is lower (with “Func” =HEAt) than [SP- |SP/5|] or higher (with
“Func” =CooL) than [SP+ |SP/5|].
- "4"    -  if  it’s  desired  to  activate  the  autotuning  automatically  to
every change of Set Point, or at the end of programmed Soft-Start
cycle.  The  Autotuning  will  start  at  the  condition  that  the  process
value is lower (with “Func” =HEAt) than [SP- |SP/5|] or higher (with
“Func” =CooL) than [SP+ |SP/5|].

Содержание TLK 33

Страница 1: ...PUTS 4 8 RS 485 SERIAL INTERFACE 4 9 PARAMETERS CONFIGURATION BY A01 5 PROGRAMMABLE PARAMETERS TABLE 6 PROBLEMS MAINTENANCE AND GUARANTEE 6 1 ERROR SIGNALLING 6 2 CLEANING 6 3 GUARANTEE AND REPAIRS 7...

Страница 2: ...an one second the value increases or decreases rapidly and after two seconds in the same condition the changing speed increases in order to allow the desired value to be reached rapidly Once the desir...

Страница 3: ...electing the desired state from the main selection menu suing the keyboard By using the key U on the keyboard suitably programming par USrb USrb tunE USrb OPLO USrb OFF it is possible to pass from rEG...

Страница 4: ...upply for each instrument as there is no isolation between the power supply and input We recommend that a check should be made that the parameters are those desired and that the application functions...

Страница 5: ...is possible to obtain a heating effect polarizing in one direction and a cooling effect when polarizing in the other direction The Double Action PID control works on the output depending on the active...

Страница 6: ...Point SP2 after a set time thus obtaining a simple automatic process cycle These functions are available for all the programmable controls PID single and double action ON OFF and Neutral Zone ON OFF...

Страница 7: ...onditions and then back in alarm conditions P V 1 0 t i m e A L 1 O N O N O N o f f o f f o f f o f f A b 1 exemple with absolute low alarm ALARM DELAY the alarm output may behave in two different way...

Страница 8: ...485 GND B HOST PC PLC 120 ohm Rt 10 8 9 10 8 9 If the instrument is equipped with a serial interface the parameters to be programmed are the following all present in the parameters group SEr Add Addre...

Страница 9: ...9 9999 0 16 rot Rotation of the measuring straight line 0 000 2 000 1 000 17 InE OPE functioning in case of measuring error Our Or Ur OUr 18 OPE Output power in case of measuring error 100 100 0 19 dI...

Страница 10: ...the correct functioning of the probe uuuu The measured variable is under the probe s limits under range oooo The measured variable is over the probe s limits over range ErAt Auto tuning not possible b...

Страница 11: ...34 MAX 12 mm 7 4 FUNCTIONAL FEATURES Control PID double action Measurement range according to the used probe see range table Display resolution according to the probe used 1 0 1 0 01 0 001 Overall acc...

Страница 12: ...a b c d e f g hh i a POWER SUPPLY G 12 24 VDC b INPUT D thermoresistances Pt100 thermocouples J K S I R mV E thermistors PTC NTC thermocouples J K S I R mV I normalized signals 0 4 20 mA V normalized...

Отзывы: