Preventative Maintenance Inspection Form
DFU-0276-4r0_fmt_en-US
PM – 4/8
Touch Screen
Initials
Action
Pass/Fail
Make sure the ablate number increases when pressing the yellow ablate "
Up
Button
" [3]. The ablate power [1] will increase in increments of 1.
Make sure the ablate number decreases when pressing the yellow ablate
"
Down Button
" [3]. The ablate power [1] will decrease in increments of 1.
Make sure the coag (coagulation) number increases when pressing the blue
coag "
Up Button
"[4]. The coag power [2] will increase in increments of 1.
Make sure the coag number decreases when pressing the blue coag "
Down
Button
"[4]. The coag power [2] will decrease in increments of 1.
Make sure you are able to press the
"Menu"
button.[5]
Once on the menu, make sure you are able to navigate through the menu
settings without any problems.
Содержание SynergyRF AR-9800
Страница 2: ...Page i This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 19 of 50 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Страница 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Страница 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Страница 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Страница 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...
Страница 55: ...SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 49 of 50 This page intentionally left blank ...