AR-9800 Display Messages
The console's Touch Screen Display [Table 2, item 5] provides information about
the status of the AR-9800 settings in real time. A list of informational and error
messages is shown in Table 4 below. Messages shaded in gray denote that
Arthrex technical support should be contacted for further instructions. Error
display messages reported to Arthrex technical services should include the Error
Message and Error Code for reference (See Figure 4, Error display).
Table 4
AR-9800 Display Messages
Message
Explanation
"Ablation Footswitch Stuck"
A button on the foot switch is stuck. Replace
the foot switch.
"COAG Footswitch Stuck"
A button on the foot switch is stuck. Replace
the foot switch.
"PROBE Already Used"
The RF device attached to the console has
either been previously used, or has been
disconnected and reconnected. Replace with a
new device.
"PROBE Communication Failure"
Communication with the RF device has been
lost. Check the connection and replace the
device, if necessary.
"Replace Probe"
The RF disposable device attached to the
console is malfunctioning.
"System Temperature Failure"
The console temperature is outside of normal
parameters, contact Arthrex technical support.
"Turn off and let cool down"
The console has overheated. Power off the
console for 10 minutes.
"WRAP Footswitch Stuck".
A button on the foot switch is stuck. Replace
the foot switch.
"Contact Technical Services"
Contact Arthrex technical services.
"Configuration File Corruption"
Contact Arthrex technical services.
"Critical Failure"
A terminal fault has occurred, contact Arthrex
technical services.
"LCD File Version Failure"
Touch panel failure, contact Arthrex technical
services.
"LCD Communications Failure"
Touch panel failure, contact Arthrex technical
services
"Main Calibration Required"
Contact Arthrex technical services.
"Power Supply Calibration
Required"
Power supply self-check failure, contact Arthrex
Technical Services.
"Power Supply Failure"
Power supply self-check failure, contact Arthrex
Содержание SynergyRF AR-9800
Страница 2: ...Page i This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 19 of 50 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Страница 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Страница 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Страница 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Страница 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...
Страница 55: ...SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 49 of 50 This page intentionally left blank ...