Preventative Maintenance Inspection Form
DFU-0276-4r0_fmt_en-US
PM – 8/8
Main Calibration Required
Contact Arthrex Technical Services
Power Supply Calibration Required
Power supply self-check failure, contact Arthrex Technical
Services.
Power Supply Failure
Power supply self-check failure, contact Arthrex Technical
Services.
Message
Possible Solution
Power Supply POST failure
Power supply self-check failure, contact Arthrex Technical
Services.
Power Supply PFC Voltage
Power supply self-check failure, contact Arthrex Technical
Services.
Power Supply Excessive Voltage
Power supply is exceeding the voltage parameters, contact
Arthrex technical services.
Power Supply Excessive Power
Power supply is exceeding the power parameters, contact
Arthrex technical services.
Relay Failure
Power supply self-check failure, contact Arthrex Technical
Services.
RF Output Failure
Power supply self-check failure, contact Arthrex Technical
Services.
Touchscreen Circuit Failure
Touch panel failure, contact Arthrex technical services.
Red Fault Message
Critical failure, Make note of message displayed. Call Arthrex
Service
Содержание SynergyRF AR-9800
Страница 2: ...Page i This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 19 of 50 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Страница 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Страница 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Страница 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Страница 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...
Страница 55: ...SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 49 of 50 This page intentionally left blank ...