Table 23 (cont.)
Guidance and Manufacturer's Statement - Electromagnetic Immunity
The AR-9800 Synergy
RF
System is intended for use in the electromagnetic environment specified
below. The customer or the user of the AR-9800 Synergy
RF
System should ensure that it is used
in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to
2.7 GHz @
1kHz AM
modulation
3 Vrms
3 V/m
Portable and mobile RF communications
equipment
should be used no closer to any part of the Model
AR-9800, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
[
]
P
V
d
1
/
5
.
3
=
= 1.2
P
[
]
P
V
d
1
/
5
.
3
=
=1.2
P
80 MHz to 800 MHz
[
]
P
E
d
1
/
7
=
= 2.3
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of
the
transmitter in watts (W) according to the
transmitter
manufacturer and
d
is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Model AR 8305 is used
Содержание Synergy AR-9800
Страница 2: ...This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 23 of 58 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Страница 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Страница 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Страница 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Страница 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...