Setup
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 33 of 58
Do not wrap Probe handpiece, Footswitch, or Generator power cord around
metal objects. Wrapping cables around metal objects may induce currents
that could lead to shock, fire or injury to patient or surgical personnel.
Only use Arthrex RF devices (Probes) that have been developed by
Arthrex specifically for the Synergy
RF
console.
Introducing the Probe into tissue without an instrument cannula may
result in tissue injury and/or product damage.
Do not insert, withdraw or touch the tip of the probe when the device
is activated.
Do not use the probe as a lever to enlarge the surgical site or gain
access to tissue.
Observe extreme caution when using electrosurgery in close
proximity to or in direct contact with metal objects or metal implants.
The majority of arthroscopes and instruments are metal. Do not
activate the probe when any portion of the metal tip is in contact with
another metal object; localized heating of the metal tip and the
adjacent metal object may result in product damage to the patient and
or operator.
The probe tip should be completely surrounded by irrigation
solution during use. A continuous flow of irrigation fluid is
recommended. Fluid flow assists in removing ablated tissue, as well
as reducing the temperature of the probe tip between activations
.
NOTES:
The disposable RF device (Probe) connector cannot be properly inserted into
the console receptacle if the corresponding black dots are not aligned.
The pins of the device connector may be damaged if the black dots are not
aligned before the device connector is fully inserted.
The user may experience intermittent operation of the disposable RF device
(Probe) if the connector is not fully seated into the console receptacle.
Insert the device connector into the console’s receptacle. Align the dots on the
connector and receptacle so they engage properly.
Содержание Synergy AR-9800
Страница 2: ...This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 23 of 58 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Страница 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Страница 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Страница 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Страница 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...