General Warnings, Training and Safety Notices - Read
This First
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 9 of 58
17.
Do NOT
clean the device with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could scratch or damage the device
18.
NEVER
allow the console receptacles to have any contact with liquids. If
there is dust or moisture on the receptacles, remove with dry compressed air.
ONLY dry connectors should be plugged into the console.
19.
Liquid on the cable connector of the accessory device can damage the device.
Before connecting the cable, ensure the receptacles are clean and dry.
20.
For aspirating Probes/disposable devices ALWAYS attach the suction tube
connector to an adequate suction source. ALWAYS ensure the suction
control valve on the device is open to ensure proper circulation of fluid.
21.
Refer to the Instructions for Use package insert Apollo
RF®
disposable devices
(DFU-0242-XX) for detailed device cleaning and sterilization instructions
included with each device Additional copies of this insert can be obtained
from the Arthrex website at
www.arthrex.com
, or by contacting your local
Arthrex Representative.
NOTE
:
This identifies information and training that can simplify the setup
and operation of this device.
1.
The user should be experienced in arthroscopic surgical techniques before using the
Arthrex Synergy
RF
System.
2.
Read this User’s Guide thoroughly before operating the device, and save it for future
reference. For additional information and training, contact your local Arthrex
representative.
3.
If required by local regulations, connect the console to the hospital equalization
connector with an equipotential cable. Connect the power cord to a wall outlet
having the correct voltage. Otherwise, product damage may result.
(The equipotential connector provided complies with IEC 60601-1:2005/A1:2012. The proper
use of equipotential cables and equipotential earthing systems must be governed by local
hospital and regulatory requirements)
4.
The AR-9800 console incorporates a universal AC input power supply. A voltage
selection switch is not required.
5.
AR-9800 has been designed to accept EMC from other devices within the limitations
as described in section 13.0.
6.
Setup for a footswitch is the same for all models. The console detects which version is
attached and allows appropriate functions.
7.
Surgeons are advised to review the product specific surgical technique prior to
performing any surgery. Arthrex provides detailed surgical techniques in print,
Содержание Synergy AR-9800
Страница 2: ...This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 23 of 58 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Страница 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Страница 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Страница 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Страница 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...