Operation and Frequently Used Functions
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 31 of 50
Accessory Handpiece Speed Limits and Deltas
The console supports individual handpiece motor-speed limits, defaults, and
speed increment and decrement “deltas”. These are defined below for the
supported handpieces.
Table 20
Accessory Handpiece Speed Limits and Delta
Forward / Reverse Speeds
(RPM)
Oscillation Speeds
(RPM)
Handpiece
Min.
Max.
Speed Delta
Min. Max.
Speed
Delta
Shaver
500
8000
300
500
3000
250
Drill
100
1400
See Drill Speeds
Section
N/A
N/A
N/A
Sag Saw
18000
18000
N/A
N/A
N/A
N/A
SJ Shaver
500
6200
300
500
3000
250
Drill Speeds
The drill supports only 100, 300, 500, 900, and 1400 (RPM). It does not cycle
through regular increments.
NOTE: While operating handpieces, the settings can be changed without risk
of damaging the motor or console.
W A R N I N G !
Do not have the device in direct contact with patient if high-frequency devices are
in use or if the patient requires defibrillation.
Console Speed Controls
The Speed Increase and Speed Decrease buttons adjust the speed. Each time a
button is pressed, the speed is adjusted as shown in Table 20. The commands are
ignored if multiple buttons are pressed simultaneously.
To make the handpiece speed increase or decrease automatically, press the Speed
Increase or Speed Decrease button for more than 1 second. The values will not
exceed the minimum and maximum values defined for the attachment.
Table 21
Handpiece Accuracy
Handpiece
ACCURACY
All Shavers ±5 % OR ±100 RPM (greater of the two)
Drill
±5 % OR ±20 RPM (greater of the two)
Sag Saw
±6 %