Setup
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 18 of 50
Try to correct the interference by following one or more of these measures:
1.
Reorient or relocate the receiving device.
2.
Increase the separation between devices.
3.
Connect the device to an outlet on a different circuit than the other
device(s) are connected.
4.
Consult the manufacturer or field service technician for receiving devices
for guidance.
Basic Setup Procedure for the AR-8305 Console
NOTE: Section
, Operation, explains how to use the console.
1.
Place the AR-8305 on a flat, dry surface, such as the AR-6481 Arthrex
Arthroscopy pump cart.
2.
Connect the receiver end of the power cord for the AR-8305 into the AC
socket and the plug end to the facility AC mains supply.
3.
Power ON the console.
4.
Allow it to fully initialize.
5.
Attach the shaver handpiece.
6.
Attach a footswitch, if applicable.
W A R N I N G !
DO NOT
stack or place equipment adjacent to the AR-8305 console, if possible.
If such a configuration is necessary, carefully observe the configuration in
question to ensure that the electromagnetic interference does not degrade the
performance.
Do not attach handpieces or footswitches during the Self Test or
the Programming Modes.
W A R N I N G !
Use only Arthrex approved accessories. Other accessories may result in
increased emissions or decreased immunity of the system. Contact your Arthrex
representative for a complete list of accessories.
DO NOT
modify any accessory.
Failure to comply may result in patient and/or operating room staff injury.
W A R N I N G !
This equipment is NOT suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen rich or nitrous oxide environment.