Product Description
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 12 of 50
Table 4
*TOOL FAILURE* Error Code Definitions
Message
Explanation
# $ @
Example: 022
Sensor 3 = Stuck Low
Sensor 2 = No Errors Found
Sensor 1 = No Errors Found
Represents Handpiece Sensor Failures
# = Sensor 3, $ = Sensor 2, @ = Sensor 1
0 = Stuck Low,
1 = Stuck High,
2 = No Errors Found.
H13
Multiple sensors shorted
H14
Multiple sensors stuck high/low
H15
Invalid Hall sensor states detected
C31
Over current 7.5A during retries
C32
Over current 5.0A during retries
C33
Over 5.0A for 6 seconds
C34
Over 7.0A for 3 seconds
C35
Over 7.5A for 1 second
C36
Over 8.0A for 250ms
C37
Over 8.1A in INT0
C38
Over 8.1A in A/D ISR
C39
Over 8.0A in FWD/REV ramp up
S51
Invalid Gear Ratio
S52
Over restart tries
S54
General system failure