Setup
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 17 of 50
W A R N I N G !
To avoid the RISK of electric shock, this equipment must only be connected to a
SUPPLY MAIN with protective earth ground.
W A R N I N G !
Do not have the device in direct contact with patient if high-frequency devices are
in use, or if the patient requires defibrillation.
Replacing the Fuses
The mains fuse is replaced with T6.3AL250v as follows:
1. Disconnect the device from the AC mains.
2.
Open the fuse tray in the AC inlet, by pulling out on the tabs.
3.
Replace the fuses with T6.3AL250v Line Fuses, as noted on the rear panel.
4. Push the fuse holder back into the AC inlet.
5. Ensure that fuse holder is fully seated and that the tabs snap back
Always use fuses with the correct values to avoid allowing
overcurrent to enter the system.
An incorrect fuse may increase the risk of electrical shock or fire
hazard.
NOTE:
The AR-8305 console incorporates a universal AC input power supply.
A voltage selection switch is not required.
Electromagnetic Compatibility
This device has passed testing for EMI / RFI radiation and
susceptibility and EMC compatibility. This device may cause
interference to other devices in the near vicinity, if not set up and
used as Arthrex instructs.
AR-8305 has been designed to accept EMC from other devices within the
limitations as described in section 14.0.
To determine if the AR-8305 is causing interference to other devices,
power OFF the AC main power switch and then ON again.