IEC 60601-1-2
Guidance and manufacturer’s declaration – electromagnetic immunity
The ArthrexID Video System is intended for use in the electromagnetic environment specified below. The customer or user of the
ArthrexIDVideo System should ensure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance Level
Electromagnetic environment - guidance
Electrostatic discharge [ESD]
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered by synthetic material the relative
humidity should be at least 30%.
Electrical fast transient / burst
IEC 61000-4-4
± 2 kV differential mode
± 1 kV for input / output
lines
100 khz Cycling
Frequency
± 2 kV differential mode
± 1 kV for input / output
lines
100 khz Cycling
Frequency
Mains power should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode.
± 2 kV common mode.
± 1 kV differential
mode.
± 2 kV common mode.
Mains power should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
U
T
= 0%,0.5 cycle
(0, 45, 90,135, 180,
225, 270, and 315
0
)
U
T
= 0%;1 cycle
U
T
= 70%;25/30 cycles
(@0 degrees)
U
T
= 70%; 250/300
cycles
U
T
= 0%,0.5 cycle
(0, 45, 90,135, 180,
225, 270, and 315
0
)
U
T
= 0%;1 cycle
U
T
= 70%;25/30 cycles
(@0 degrees)
U
T
= 70%; 250/300
cycles
Mains power should be that of a typical commercial
or hospital environment. If the user of the ArthrexID
Video System requires continued operation during
power mains interruptions, it is recommended that
the ArthrexID Video System be powered from an
uninterruptible power supply.
If voltage dips or short interruptions do occur the
CCU will recover in < 5 seconds but the LED Light
Source may be in stand-by. User will need to turn
on the LED Light Engine with the front panel button.
Power Frequency [50/60 Hz]
magnetic field.
IEC 61000-4-8
3 A/m
3 A/m @50 & 60 Hz
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.