15A3356O
APPENDIX B
PCS PRINTOUTS
Sample PCS Instrument Calibration Check Printout
(Your Lab's Name)
Date: 01 Jun 2011
Time: 15:03
PCS INSTRUMENT CALIBRATION CHECK
Operator ID: 39
Cal Kit Lot Code: 39341
READING DATA
============
520 nm 520 nm 730 nm 730 nm
Cal Avg Abs Std Dev Avg Abs Std Dev
A 0.0000 0.00005 0.0000 0.00002
B 0.5019 0.00006 0.4959 0.00002
C 0.9926 0.00008 0.9878 0.00004
D 1.4789 0.00004 1.4798 0.00006
INSTRUMENT ACCURACY
====================
520/730 Cal Kit %
Cal Ratio Ratio Error Result
B 1.0121 1.0122 0.00% Pass
C 1.0049 1.0050 -0.01% Pass
D 0.9994 0.9985 0.09% Pass
Spec. (all 3 cases): +/- 0.75%
INSTRUMENT REPEATABILITY
========================
520 nm:
Cal Std Dev Spec. Result
A 0.00005 0.00018 Pass
B 0.00006 0.00016 Pass
C 0.00008 0.00023 Pass
D 0.00004 0.00050 Pass
730 nm:
Cal Std Dev Spec. Result
A 0.00002 0.00050 Pass
B 0.00002 0.00050 Pass
C 0.00004 0.00050 Pass
D 0.00006 0.00050 Pass
OVERALL CALIBRATION RESULT: PASS
Temperature: 21.2 C
Instrument Cal. Number: 105
Instrument Serial No.: 8282
Software Version: PCS 7A2.203
ARTEL PCS PIPETTE CALIBRATION SYSTEM
Patents:
USA: 5492673, 5298978, 7772008, 7791716
Europe: EP0711404, EP0628157
Japan: 2520852
Содержание PCS 2
Страница 6: ...15A3356O 2 This Page Intentionally Left Blank ...
Страница 10: ...15A3356O 6 This Page Intentionally Left Blank ...
Страница 48: ...15A3356O 44 This Page Intentionally Left Blank ...
Страница 52: ...15A3356O 48 This Page Intentionally Left Blank ...
Страница 54: ...15A3356O This Page Intentionally Left Blank ...
Страница 60: ...15A3356O This Page Intentionally Left Blank ...
Страница 62: ...15A3356O This Page Intentionally Left Blank ...
Страница 64: ...15A3356O This Page Intentionally Left Blank ...
Страница 66: ...15A3356O This Page Intentionally Left Blank ...
Страница 68: ...15A3356O This Page Intentionally Left Blank ...