AE-780 Operation Manual
Rev. 11/30/12 MKB
AE-780 C
ALIBRATION
I
NSTRUCTIONS
This meter should be calibrated every twelve months. To verify if the tester is within
specification, a test resistance can be applied across the parallel bars using a
resistance decade box. Calibration can be obtained by contacting Armeka Engineering
Ltd..
S
PECIFICATIONS
Dimensions:
70mm x 130mm x 35mm (approximate)
Weight:
103 grams (approximate)
Power:
Battery
-
operated PP3 9
-
volt
Connections:
2 x 3.5mm jack plug for earth connection
Test Range:
10
3
to 10
12+
Dissipative Range:
3 x 10
5
to 3 x 10
9
(½ decade between each decade on a logarithmic scale)
Unit of Measurement:
Surface resistivity
Ohms per square
Resistance point to point
Ohms
Accuracy:
+/
‐
0.5 decade in conductive range
+/
‐
0.25 decade in dissipative range
Decade range
Change over point
value
value
10E3 to 4
3x10E3
3,000
3 K ohm
10E4 to 5
3x10E4
30,000
30 K ohm
10E5 to 3x10E5
1.7x10E5
170,000
170 K ohm
3x10E5 to 10E6
6x10E5
600,000
600 K ohm
10E6 to 3x10E6
1.7x10E6
1,700,000
1.7 M ohm
3x10E6 to 10E7
6x10E6
6,000,000
6 M ohm
10E7 to 3x10E7
1.7x10E7
17,000,000
17 M ohm
3x10E7 to 10E8
6x10E7
60,000,000
60 M ohm
10E8 to 3x10E8
1.7x10E8
170,000,000
170 M ohm
3x10E8 to 10E9
6x10E8
600,000,000
600 M ohm
10E9 to 3x10E9
1.7x10E9
1,700,000,000
1.7 G ohm
3x10E9 to 10E10
6x10E9
6,000,000,000
6 G ohm
10E10 to 11
3x10E10
30,000,000,000
30 G ohm
Eastern Energy Co., Ltd.
Tel: 66-3844-6117
40/4 Vitoondumri Rd., Banbueng
Page 12 of 13
Banbueng, Chonburi, Thailand
www.eastern-energy.com
www.eastern-energy.net