background image

 

AE-780 Operation Manual

 

Rev. 11/30/12 MKB

 

 
T

EST

 V

OLTAGE

 

 

The test voltages are 10v and 100v. According to ESD Association (ESDA) standards 
S4.1 and S7.1, 10v should be applied to surfaces with resistivity of less than 10

6

 and 

100v should be applied to surfaces with resistivity of 10

6

 or greater. The AE-780 will 

automatically simulate the proper voltage according to the test measurement.

 

 

As defined by the ESDA, values indicate the following:

 

 

 

Voltage

 

Range

Definition

 

10 volt

< 10

6

 ohms per square

Conductive

 

100 volt

 

10

6

 – 10

11

 ohms per square

Dissipative

 

100 volt

 

> 10

11

 ohms per square

Insulative

A N

OTE ABOUT

 V

OLTAGE

   

 

 

In previous years, those desiring to measure resistivity or resistance followed the ASTM 
D264, ASTM 991, NFPA 56A or NFPA 99 test standards. These procedures required 
one to test at either 500 or 1000 volts. This caused concern regarding safety of the 
person performing the tests. The ESDA standardized the test procedures so that lower 
volts could be used within specific ranges.

 

 

The AE-780 meter uses a 9

-

volt battery. Some meters with 9

-

volt batteries do not give 

the accuracy needed to perform the tests, especially at values higher than 10

7

. The AE-

780 is built with a transformer that converts the 9

-

volt charge from the battery to 10 

volts or 100 volts. The meter applies a constant charge over the complete voltage 
range. Accuracy depends on applied voltage, temperature, and humidity.

 

 

T

EMPERATURE AND

 H

UMIDITY

 

 

Ambient humidity and temperature affect the electrical properties of the material being 
tested. The combination of low humidity and low temperature will give the highest 
electrical resistance results and slowest dissipation times. At high humidity, a thin layer 
of water is condensed on or absorbed into the material being tested. This is true of 
hydroscopic additives that are incorporated into a material in order to increase the 
electrical conductivity. These additives will allow moisture to be absorbed into the 
materials to which they are added.

 

 

At elevated temperatures, the mobility of free electrons increases, thereby also 
increasing a material’s conductivity. This is especially true for carbon black, metallic 
oxides, metals, and other substances. When the material is at a lower temperature, 
built

in stresses occur which may increase resistance due to increased distance 

between the conductive additives. Thus, humidity and temperature must be known.

 

Eastern Energy Co., Ltd.

 

Tel: 66-3844-6117

40/4 Vitoondumri Rd., Banbueng

   

Page 4  of 13

[email protected]

Banbueng, Chonburi, Thailand

 

 

www.eastern-energy.com

www.eastern-energy.net

Содержание AE-780

Страница 1: ...N MANUAL Meter is warranted for one year from the date of purchase on parts and labor Calibration is recommended every 12 months Eastern Energy Co Ltd Tel 66 3844 6117 40 4 Vitoondumri Rd Banbueng Page 1 of 13 sale ete co th Banbueng Chonburi Thailand www eastern energy com ...

Страница 2: ...ith the 5 lb probes available in the accessory kit it will comply to IEC 61340 4 1 ANSI ESDA S4 1 and ANSI ESDA S7 1 AE 780 Resistance Meter includes Tester Two accordion cables stereo to banana 9 volt battery Certificate of calibration AE 780 Accessory Kit includes Two 5 lb probes Foam lined carrying case Limits Resistivity 103 1012 ohms square Resistance 103 1012 ohms Measuring voltage 10v and 1...

Страница 3: ...ter s test values for resistance are in ohms The test value is indicated on the LED display Half decades provide greater accuracy by giving a closer approximation to the measurement value An LED will brighten to the according test result Colors signify the test value s function COLOR INDICATING FUNCTION Ω Green Conductive 103 105 Yellow Dissipative ideal test measurement 3 x 105 109 Orange Dissipa...

Страница 4: ...t with a transformer that converts the 9 volt charge from the battery to 10 volts or 100 volts The meter applies a constant charge over the complete voltage range Accuracy depends on applied voltage temperature and humidity TEMPERATURE AND HUMIDITY Ambient humidity and temperature affect the electrical properties of the material being tested The combination of low humidity and low temperature will...

Страница 5: ...ith a certificate of calibration which verifies the calibration of the instrument using equipment that is traceable to National Standards and CAD generated techniques Meters can be sent back to Armeka Engineering Ltd for calibration for a lab fee or they can be sent to a certified calibration lab See page 12 for calibration instructions The meters also come with a CE mark approval MEASURING WITH I...

Страница 6: ... in ohms per square although they are displayed in ohms F The test value is indicated on the LED display A decade will brighten to the according test value Colors signify the test value s function see page 3 MEASURING WITH EXTERNAL PROBES When the measurement is taken using the 5 lb external probes from the AE 780 accessory kit the tester will indicate the resistance of the material in ohms The ex...

Страница 7: ... preparation probe preparation and spacing of the 5 lb probes Select and read the correct test procedure or standard for the desired measurement Test procedures can be purchased from www eastern energy com A Connect one end of each banana test lead into the sockets of the meter Connect the other end of the test coil cords into the 5 lb probes B Place both probes on the material according to test p...

Страница 8: ...ain an average measurement of a floor map out a 4 x 4 section and conduct five tests one at a time within the square Conduct a test for each side of the square and a final test diagonally through the center as shown in the drawing below In each RTT test the 5 lb probes are placed three feet apart 36 inches Connect the test leads to the meter Attach a 5 lb probe to the end of each lead and place th...

Страница 9: ...e RTG tests per 5 000 sq ft Connect the leads for the external probes to the meter When the cables have been plugged into the appropriate socket the parallel probes under the meter disengage Attach one lead to a 5 lb probe and place probe onto the floor to be tested Attach the other lead to an alligator clip and connect to a groundable point RTGP If using a ground adapter plug plug the banana lead...

Страница 10: ...und snap or connect the banana plug to a common point ground plug b Press the TEST button and the value will be displayed on the LED While performing test do not touch lead wires or probe Avoid overlapping of lead wires This will ensure accurate readings c Resistance values are in ohms When recording test values also make note of the temperature and humidity as the environment can affect test resu...

Страница 11: ...a conductive metal plate such as stainless steel Place one of the 5 lb probes onto the material so that the material is sandwiched between the probe and metal plate See below C Place the second 5 lb probe on the conductive metal plate D Press the TEST button and the value will be displayed on the LED Volume Resistance is measured in ohms cm Eastern Energy Co Ltd Tel 66 3844 6117 40 4 Vitoondumri R...

Страница 12: ...quare Resistance point to point Ohms Accuracy 0 5 decade in conductive range 0 25 decade in dissipative range Decade range Change over point value value 10E3 to 4 3x10E3 3 000 3 K ohm 10E4 to 5 3x10E4 30 000 30 K ohm 10E5 to 3x10E5 1 7x10E5 170 000 170 K ohm 3x10E5 to 10E6 6x10E5 600 000 600 K ohm 10E6 to 3x10E6 1 7x10E6 1 700 000 1 7 M ohm 3x10E6 to 10E7 6x10E6 6 000 000 6 M ohm 10E7 to 3x10E7 1 ...

Страница 13: ...ades increase the resistance of the decade box while pressing the meter s test button Record the resistance when the next LED lights permanently this is the changeover resistance Example The first green LED is illuminated 103 1 kilohm At 3 or 4 kilohms 104 LED is illuminated The changeover point is 3 or 4 kilohms 104 10 kilohms so between 3 or 4 kilohms and 30 or 40 kilohms will be the changeover ...

Отзывы: