INTRODUCTION TO OPTIONS
This section will cover the operation and setup of all of the options available for your scale. It is important to note that some described
menu items will be missing if that option was not installed in your specific scale. All of the options listed may be installed in any scale
and may run concurrently. Any of the options may be installed later at the factory if it wasn’t installed at time of purchase
. Call our
service department for details.
RS422 COMMUNICATION
The RS422 option is a fully capable, bi-directional communications port. The port can be configured to operate at a variety of baud
rates and the output data frames are definable by the user. The printing of a frame can be initiated by pressing the print button, by an
external command or continuously when the print stream mode is activated.
The communications port also contains an extensive external command interface allowing the ability to perform remote key presses.
RS422
A
B
Z
Y
GND
Configuring the Port
Baud rate and other parameters can be configured through the Port Configuration screen that can be accessed under menu SETUP
MENU/OPTION SETUPS/RS485/PORT CONFIG. The options are shown below. The UI
of “RS485” is used for communication via RS422.
BAUD RATE
Settable to 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200, 31250 or 38400 bits per second.
DATA BITS
Seven or Eight
STOP BITS
One or Two
PARITY
Even, odd or none
ECHO
On or off. When echo is enabled the scale will echo each character the scale receives back to the user.
<RS422 indicator depiction>
Perform Test Print
As an aid to port configuration we have included a “perform test print” function to the RS
485 menu. Executing this will cause your
current Print Frame to print. To access this function, go to the menu SETUP MENU/OPTION SETUPS/RS485/PERFORM TEST PRINT.
Loop Back Test
You may also test the operation of the RS422 port by performing a loop back test. On the output connector, A->Z and B->Y
(transmitted and received data) need to be shorted together. When you perform the test, the scale sends data and reads it back
through its own port, where it is compared and the results are show.