Important Selections
The two most important selections are:
❑
Span Calibration
❑
SAW Correction
Span Calibration
Span calibration adjusts the platform’s sensitivity so that the display reads correctly. A calibrated weight is required to p
erform this
and the procedure is outlined above in the
SAW Scale Full Calibration
section.
SAW Correction
This procedure corrects parametric deviances that naturally occurs over time with the SAW load cell. Whenever the scale starts to
show some odd behavior (such as weights not reading accurately, or slight jumping), this should be the first procedure to try to
mitigate these issues.
Other Items
Other selections on this menu are:
❑
Edit Description
❑
SAW Diagnostics
❑
*Resolution-Overload*
❑
CAP Zero
❑
Zero Tracking
❑
Stability Control
❑
Software Filter
❑
Zero/Motion Detect
❑
Auto Zero
❑
*Platform Auto Setup*
❑
Load Cell ID#
*These items are for Advanced Use only. Accessing and changing them in any way without proper supervision would adversely affect
your scale.
Parametric Settings
Edit
Description
Allows you to edit fourteen-character description can be assigned to the p
latform. The default description is “P1”. This should not be
changed for any reason unless you want to the description as part of your Print Frame. However, changing the description will not
affect the metrological properties of your scale.
Resolution-Overload
This setting should not be changed unless instructed by a Service Technician.
The scale has the capability of displaying its reading in any of eight standard conversion units. Only four are enabled by default and the
user can enable the rest. The conversion units that are automatically enabled are pounds, kilograms, grams and ounces. The remaining
conversions are troy ounces, pennyweights, grains and a user defined conversion.
Each conversion unit’s resolution can be set individually. To do this, select RESOLUTION
-OVERLOAD from the platform menu. A list of
conversions is displayed and each active conversion will have a check mark to its right.