Advanced Setup and Diagnostics
SAW Status
This screen checks all the sensors in the SAW platform to make sure they are operating within a certain range. If the parameter
readings are all within expected range, the sensor status will show “OK”. Otherwise, it will show as “Fail”.
SAW Diagnostics
Shows the digital raw parameters of all the sensors equipped within the scale. This screen is primarily used for service and
troubleshooting.
Platform Auto Setup
This setting should not be accessed unless instructed by a Service Technician.
Auto Setup will
erase
all of the
parameters for the selected platform and return it to the default state. It also erases the platform’s
span and cornering calibration. Auto setup is useful when the scale is first manufactured or when the user inadvertently changes a
calibration parameter that adversely affects the scale and then forgets the calibration parameter or its original value. Upon activation,
you will be presented with a selection of platform capacities and resolutions. Pick the appropriate entry and then press ENTER. Auto
setup will be performed. A span calibration is now necessary.