MAGX2 User Guide
52
V2.14 1-2-2017
6.
Maintenance
MAGX2 flowmeter does not require any special maintenance. Dependent on the media being measured it is
recommended that approximately once a year, to remove the sensor from the pipe and clean the liner. Method
of cleaning consists of removing mechanical dirt and any non-conductive coating (like oil film) from the liner. A
very dirty liner could cause inaccuracy of the measurement. Check mechanical state of the liner.
6.1. Self
–cleaning electrodes
If mechanical cleaning is not possible, MAGX2 has electrolytic method to clean electrodes.
An electrolytic method is advantageous for its simplicity, however it can only be applied for the contamination
that can be removed by electrolysis. (Low contamination and deposit).
24VAC voltage is applied directly to sensor electrodes to clean them. The time that that voltage is applied is
selectable for user from 1 to 9999 seconds. For more info please go to section 3.5 User settings.
GPRS
BLUETOOTH
TCP/IP
RS232
RS485
USB