MAGX2 User Guide
21
V2.14 1-2-2017
3.5.12. User Settings > Current Loop
Back with no change
Item selection
Enter this part of the menu
User Settings > Current Loop > Settings Signal
This function serves to select which signal the output should be
giving.
Back with no change
Item selection
Selection identification
Confirmation of setup and saving to memory
Flow+
Output:
10 mA
, for any positive flow
Flow
–
Output:
10 mA
, for any negative flow
Error
Output:
10 mA
, for any error identified by the device
The signal can be cancelled by pressing any
push button on the flowmeter.
Air Detect
Output:
10 mA
, during air detection (empty pipe)
Fixed
Output: fixed output of 10 mA
Direct
Driving
Output: Direct Driving
– setup is below
OFF
Output: fixed output of
4 mA
User Settings > Current Loop > Direct Driving
This function serves to set flow values in relation to current output.
Possible range:
0
– 36000 m3/h
;
4
– 20 mA
Back with no change
Selection of digit position
Value setting
Confirmation of setup and saving to memory
Flow
min.
– max.
Setup of measurement flow-range (only positive
values)
Current
min.
– max.
Setup of the current output range, corresponding to
the actual flow-rate within the given range
When changing an item for the current
loop output, all settings for the voltage
output are to be changed, to make sure
there will be no signal conflict at the
output port. The output that is not used has to be
switched off (Settings
– Signal – OFF).