MAGE1 User Guide
24
V1.5 12-1-2022
Delete Total Volume
–
write value different to zero for erasing the Total flow totalizer.
Delete Batch
–
write value different to zero for erasing the Total
– flow totalizer.
Delete Total + Volume
–
write value different to zero for erasing the Total + flow totalizer
Delete OK (min)
–
write value different to zero for erasing the OK min counter.
Delete Error (min)
–
write value different to zero for erasing the Error min counter
Flow Simulation
– switch off/on the simulation flow function. Default OFF
Simulated Flow
– write simulated flow. Real value = Actual value / 1000
Diameter
– diameter of the sensor.
Unit No.
– the serial number of unit
Flow Qn
– setup to the excepted flow Qn. It is set automatically when you write diameter.
Real value = Actual value / 1000
Calibration Point 1
– calibration flow 1
Calibration Point 2
– calibration flow 2
Calibration Point 3
– calibration flow 3
Calibration Data 1
– calibration data 1
Calibration Data 2
– calibration data 2
Calibration Data 3
– calibration data 3
Note: CP1 < CP3 < CP2 and CD1 < CD3 < CD2.
The calibration point 3 and the calibration data 3 could be set to zero. CP1 < CP2 and CD1 < CD3.
Zero Flow Set
–
after activation this function, next 125 samples are compute to average value for zero flow
constant
Zero Flow Erase
– erase zero flow constant to 0
Zero Flow Constant
– Set manually value for zero flow constant
Excitation Frequency
– choose the excitation frequency.
Excitation
– write zero for turn OFF the excitation. Default ON
Reset
– Reset all system. After application it must be restarted by connecting SW with flowmeter
CP1
CP2
CD2
CD1
Flow
ADC data
CP3
CD3
CD1 = Calibration data 1
CD2 = Calibration data 2
CD3 = Calibration data 3
CP1 = Calibration point 1
CP2 = Calibration point 2
CP3 = Calibration point 3
ADC data = data from AD
converter