MAGE1 User Guide
34
V1.5 12-1-2022
13.2. Trouble: Non stable flow or Empty pipe alarm
Is flowmeter
installed according
to the installation
manual
Make correct
installation
No
Check if there is any
source of electrical
noise near the
flowmeter or pipe
Turn off all sources
of noise and check if
the flow is stable
now
Yes
Is the pipe totally full
with measured liquid
all the time?
Eliminate
the
electrical
noise
Empty pipe is confirmed.
Make better installation of
the flowmeter
Has the liquid
minimal conductivity
≥ 20µS/cm
Electromagnetic
flowmeter is not
applicable for this liquid
Contact technical support
Are there bubbles in
the liquid?
Eliminate the air bubbles.
No
No
No
No
Yes
Yes
Yes
Yes
Yes
No