Care and Maintenance
Preventive Maintenance Schedule
The equipment is subjected to wear and tear, and the following maintenance instructions must be acted upon
when specified to ensure that the equipment remains within its original manufacturing specifications. Care and
maintenance must be carried out in accordance with the preventive maintenance schedule below.
Customer obligations must be carried out by qualified personnel in accordance with the instructions in this
manual.
User Inspections
WARNING: The maintenance described in the following checklist is the minimum that the
manufacturer recommends. In some cases more frequent inspections should be carried out.
Continuing to use this equipment without conducting regular inspections or when a fault is
found will seriously compromise the safety of the user and of the resident. Local regulations and
standards may be higher than those of the manufacturer. A load test for the ceiling lift is
recommended. Preventive maintenance specified in this manual can prevent accidents and
reduce repair costs.
WARNING: Safety related maintenance and authorized service must be carried out by qualified
personnel, fully trained in servicing procedures by ArjoHuntleigh and equipped with proper
tools. Failure to meet these requirements could result in personal injuries and/or unsafe
equipment.
FREQUENCY
Inspections for lift
and track system
Initially
Before
every
use
Every two
months or
500 cycles
Every four
months or
1000 cycles
Every year
or
2500 cycles
Every two
years or
5000 cycles
Inspect for evidence of
external damage, missing
parts or broken panels.
X
X
Make sure that end
stoppers and rail caps are
in place and tightened.
X
X
Inspect strap for wear,
discoloration or loose
threads.
X
Recharge batteries.
X
Inspect wheels in rail for
damage, rust or cracks.
Replace if damaged.
X
Clean the rail and the clip-
on charging station
contacts.
X
Overall inspection by
authorized personnel.
X
Verify emergency stop
cord.
X
Verify emergency lowering
device.
X